Department of Restorative Dentistry, Faculty of Dentistry, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, Minas Gerais, Brazil.
J Endod. 2011 Aug;37(8):1124-7. doi: 10.1016/j.joen.2011.05.010. Epub 2011 Jun 25.
The purpose of this study was to evaluate the torsional behavior of rotary ProTaper Universal instruments after multiple clinical uses with and without lateral pressure motion.
Thirty sets of ProTaper Universal instruments were divided into 3 groups (n = 10): control group (CG) involving new instruments tested in torsion on the basis of ISO 3630-1, lateral pressure (LP) group involving instruments that were clinically used on 5 molars by using the shaping instruments S1 and S2 with lateral pressure motion, and no lateral pressure (NLP) group involving instruments that were clinically used on 5 molars without lateral pressure motion. The instruments in the LP and NLP groups were subsequently tested in torsion. Data were analyzed by using analysis of variance (α = 0.05).
Multiple clinical uses caused a reduction in the maximum torque in the analyzed instruments. When the effect of using lateral pressure motion with the shaping instruments was assessed, a tendency of reduction in the maximum torque for the S1 and S2 instruments and of increase for the F1 and F2 instruments was observed in the group with lateral pressure motion.
The use of lateral pressure motion with the shaping instruments S1 and S2 tended to produce smaller decreases in the torsional resistance of the finishing instruments F1 and F2.
本研究旨在评估 ProTaper Universal 机用镍钛器械在多次临床使用后有无侧向加压运动时的扭转行为。
将 30 套 ProTaper Universal 器械分为 3 组(n = 10):对照组(CG),即新器械根据 ISO 3630-1 在扭转试验中进行测试;侧向加压组(LP),即用 S1 和 S2 成形器械施加侧向压力对 5 颗磨牙进行临床使用的器械;无侧向加压组(NLP),即用无侧向压力运动的器械对 5 颗磨牙进行临床使用的器械。LP 和 NLP 组的器械随后进行扭转试验。采用方差分析(α = 0.05)对数据进行分析。
多次临床使用导致分析器械的最大扭矩降低。当评估使用成形器械施加侧向压力运动的效果时,在施加侧向压力运动的组中,S1 和 S2 器械的最大扭矩呈降低趋势,而 F1 和 F2 器械的最大扭矩呈升高趋势。
使用 S1 和 S2 成形器械施加侧向压力运动,可能会减小终末器械 F1 和 F2 的扭转阻力降低幅度。