The aim of this study was to evaluate clinical and radiographic outcomes of Elektra® trapeziometacarpal prostheses after 2 years. We present a longitudinal cohort study of 19 prostheses for the treatment of Eaton stage II and III osteoarthritis (mean follow-up of 29 months). QuickDASH score, pinch strength and mobility were determined, and radiographs were analysed. Isotope scans were taken in patients with persistent pain or suspected loosening. Although the QuickDASH score was 69 before and 38 after surgery, nine patients had pain at the trapeziometacarpal joint with radiographic osteolysis around the trapezium component and a positive bone scan uptake at subsequent follow-up. The implant was revised in three of these nine patients and another patient underwent surgery for periprosthetic fracture. Only ten of the 19 implants showed no sign of failure. Most problems derived from the trapezium cup. Because of these poor outcomes after only 2 years, we cannot recommend this implant.
本研究的目的是评估 Elektra® 舟大多角骨假体在 2 年后的临床和影像学结果。我们开展了一项纵向队列研究,纳入 19 个用于治疗伊顿 II 期和 III 期骨关节炎的假体(平均随访 29 个月)。测定了 QuickDASH 评分、捏力和活动度,并对 X 光片进行了分析。对持续疼痛或怀疑假体松动的患者进行了同位素扫描。尽管术前 QuickDASH 评分为 69 分,术后为 38 分,但 9 例患者在舟大多角骨关节处疼痛,大多角骨组件周围出现影像学骨质溶解,且在后续随访中骨扫描摄取呈阳性。这 9 例患者中有 3 例进行了假体翻修,另 1 例患者因假体周围骨折接受了手术。19 个植入物中只有 10 个没有失败迹象。大多数问题源于大多角骨杯。由于仅 2 年后就出现了这些不良结果,我们不推荐使用这种植入物。