Ninewells Hospital, Dundee, UK.
Foot Ankle Surg. 2011 Sep;17(3):124-7. doi: 10.1016/j.fas.2010.02.006. Epub 2010 May 23.
Variations in the relative lengths of the metatarsal bones have been implicated as a cause of forefoot pain. Osteotomies to balance the metatarsal lengths have been described as therapy. Although measurement of metatarsal length is required for these osteotomies, there is no gold standard method to measure it radiologically. We compared the three main radiological methods to measure the second metatarsal length described in the literature.
Standing dorsoplantar radiographs of 81 feet were measured using the methods described by Coughlin, Maestro and Hardy and Clapham. Bland and Altman's method was used to compare the above techniques for reproducibility and agreement between each method.
The three methods produced markedly different results. Coughlin's and Maestro's methods showed a relative protrusion of the second metatarsal-relatively longer than the first metatarsal by an average of 4 & 3 mm respectively. Hardy and Clapham's method, however, showed a relative retraction of second metatarsal by an average of 1 mm. Both intra-observer and inter-observer errors were smallest for the Coughlin method (the most reproducible) and largest for the Hardy/Clapham method. Significant lack of agreement was found when we compared the 3 techniques with bias for measuring the mean between methods ranging from 1mm to 5.5mm.
Metatarsal lengths as measured by these three methods can be very variable. Thus their role in planning metatarsal osteotomies and establishing relationship of metatarsal protrusion with metatarsophalangeal joint instability and other similar conditions is questionable and require further studies.
跖骨相对长度的变化被认为是导致前脚疼痛的原因。为了平衡跖骨长度,已经描述了截骨术作为一种治疗方法。尽管这些截骨术需要测量跖骨长度,但没有放射学上的金标准方法来测量它。我们比较了文献中描述的三种主要的测量第二跖骨长度的放射学方法。
对 81 只脚的站立前后位 X 线片,分别采用 Coughlin、Maestro 和 Hardy 与 Clapham 描述的方法进行测量。使用 Bland 和 Altman 的方法比较了上述三种技术的可重复性和每种方法之间的一致性。
这三种方法得出的结果明显不同。Coughlin 和 Maestro 的方法显示第二跖骨相对突出,比第一跖骨长平均 4 毫米和 3 毫米。然而,Hardy 和 Clapham 的方法显示第二跖骨相对回缩,平均为 1 毫米。观察者内和观察者间误差最小的是 Coughlin 方法(最具可重复性),最大的是 Hardy/Clapham 方法。当我们比较这三种技术时,发现存在显著的缺乏一致性,测量方法之间的平均偏差为 1 毫米至 5.5 毫米。
这三种方法测量的跖骨长度可能有很大的差异。因此,它们在计划跖骨截骨术和确定跖骨突出与跖趾关节不稳定和其他类似情况之间的关系中的作用是值得怀疑的,需要进一步的研究。