Noninvasive measurement of local arterial pulse wave velocity in humans by ultrasound.

作者信息

Zhou Lixia, Beach Kirk W, Liang Hai-Dong, Halliwell Mike, Wells Peter N T

机构信息

University Hospitals Bristol NHS Foundation Trust, Department of Medical Physics and Bioengineering, Bristol General Hospital, Bristol BS1 6SY, UK.

出版信息

J Med Eng Technol. 2011 Aug-Oct;35(6-7):362-9. doi: 10.3109/03091902.2011.595532. Epub 2011 Jul 27.

Abstract

An improved method for noninvasive measurement of the local velocity of arterial pulse wave propagation by an echo-tracking-based ultrasound system is described. A data acquisition image interface was programmed in the ultrasound machine simultaneously to record M-mode ultrasound signals at two locations of a given distance apart along an artery. The selections of measurement sites, separation, and time resolution were performed on the control interface. The temporal sampling frequency could be as high as 10 kHz. The displacements of the blood vessel wall along the time axis were calculated from the M-mode signals by cross-correlation of the radio-frequency data and the distension waveforms were obtained. The temporal separation of the feet of the distension curves from the two measurement locations was derived to give the travel time of the pulse wave. Measurements were made in vivo on human carotid arteries. The pulse wave velocities measured from four volunteers were from 4.1 to 7.2 m/s with coefficients of variation from 5.9 to 29.5%. Some of the factors contributing to the variation in measured values of the velocity are discussed. The method is simple to implement and should be suitable for clinical research into local pulse wave velocity.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验