Institute of General and Physical Chemistry, Belgrade, Serbia.
Talanta. 2011 Sep 15;85(3):1398-403. doi: 10.1016/j.talanta.2011.06.039. Epub 2011 Jun 17.
Behavior of hydrogen peroxide in alkaline medium has been studied by direct current (DC) polarography with dropping mercury electrode (DME) aiming to apply it in antioxidant (AO) activity determination. Development of a peroxide anodic current having form of a peak, instead of common polarographic wave, has been investigated. As a base for this investigation the interaction of H(2)O(2) with anodically dissolved mercury was followed. Formation of mercury complex [Hg(O(2)H)(OH)] has been confirmed. The relevant experimental conditions, such as temperature, concentration and pH dependence, as well as time stability of hydrogen peroxide anodic current, have been assessed. Development of an AO assay based on decrease of anodic current of hydrogen peroxide in the presence of antioxidants (AOs) has been described. Under optimized working conditions, a series of benzoic acids along with corresponding cinnamate analogues have been tested for hydrogen peroxide scavenging activity. In addition, the assay versatility has been confirmed on various complex samples.
采用直流极谱法和滴汞电极(DME)研究了过氧化氢在碱性介质中的行为,旨在将其应用于抗氧化剂(AO)活性的测定。研究了过氧阳极电流的发展,其形式为峰,而不是常见的极谱波。作为这项研究的基础,我们跟踪了 H(2)O(2)与阳极溶解汞的相互作用。已经证实形成了汞配合物[Hg(O(2)H)(OH)]。评估了相关的实验条件,如温度、浓度和 pH 依赖性,以及过氧化氢阳极电流的时间稳定性。描述了一种基于抗氧化剂(AOs)存在下过氧化氢阳极电流减少的 AO 测定法。在优化的工作条件下,对一系列苯甲酸及其相应肉桂酸类似物进行了过氧化氢清除活性测试。此外,还在各种复杂样品上证实了该测定法的多功能性。