Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Niigata University, Niigata 951-8518, Japan.
Med Phys. 2011 Jun;38(6):2850-8. doi: 10.1118/1.3584200.
This paper describes the low-kV spectral optimization of dual-energy CT (DECT) equipped with high-kV tin filtration for the quantitative acquisition of electron density information, which is essential for treatment planning in radiotherapy. In addition, an analytical DECT image simulation was preliminarily performed to demonstrate the effectiveness of the optimized DECT with respect to the beam-hardening reduction.
To optimize the low-kV spectrum of DECT, the author calculated the beam-hardening error, CT dose index, and tube loadings for a 50-cm diameter cylindrical water phantom with various combinations of filter materials, a range of thicknesses, and low-kV tube potentials. In addition, a single tube potential of 140 kV filtered by 0.4 mm tin (Sn) was employed for high-kV scanning, as is similar to the commercial implementation of the second-generation dual-source CT scanner. The optimized spectral parameters were then applied to the analytical DECT image simulation using two-dimensional fan-beam geometry for a virtual solid water phantom with 16 bodylike tissue inserts.
The author predicts that an optimal low-kV filtration would be 0.144-mm tungsten (W) at 90 kV, as it yields a minimal beam-hardening error with lower tube loadings and dose. The high-kV tube loading and dose obtained using the W filtration were 99 mAs and 2.2 mGy, respectively. These values are nearly equal to those obtained in the case of 2.5 mm Al at 100 kV (100 mAs and 2.3 mGy), which was regarded in this study as a reference filtration; however, the W filtration significantly reduced the beam-hardening error, from 9.5 to 1.4%. The corresponding low-kV tube loading (112 mAs) was five times greater than that of the reference case (21 mAs), but it was maintained at a certain practical level since the low-kV tube loading was comparable to the high-kV tube loading of the reference (100 mAs). The superiority of the beam-hardening reduction is reflected in the simulated images; for example, by the use of the W filter, the beam-hardening-induced deviation between the simulated and theoretical electron density values of cortical bone was reduced from 7.4 to 1.2% as compared with the reference filtration, even though no correction for beam hardening was performed.
In terms of beam hardening reduction, the DECT with the low-kV W filtration is more effective for the quantitative measurement of electron density within a practical limit of tube loadings and without additional dose.
本文描述了配备高千伏锡滤器的双能 CT(DECT)的低千伏能谱优化,用于定量获取电子密度信息,这对于放射治疗中的治疗计划至关重要。此外,还初步进行了分析性 DECT 图像模拟,以证明优化后的 DECT 在减少束硬化方面的有效性。
为了优化 DECT 的低千伏谱,作者计算了不同过滤材料、厚度范围和低千伏管电压组合的 50cm 直径圆柱形水模的束硬化误差、CT 剂量指数和管负载。此外,还采用 0.4mm 锡(Sn)过滤的单管电位 140kV 进行高千伏扫描,类似于第二代双源 CT 扫描仪的商业实现。然后,将优化的光谱参数应用于二维扇形束几何形状的分析性 DECT 图像模拟,用于具有 16 个体状组织插入物的虚拟实心水模。
作者预测,最佳低千伏过滤将是 0.144mm 钨(W)在 90kV,因为它具有最低的束硬化误差和更低的管负载和剂量。使用 W 过滤获得的高千伏管负载和剂量分别为 99mAs 和 2.2mGy,这几乎与在 100kV 下使用 2.5mmAl 获得的值(100mAs 和 2.3mGy)相等;然而,W 过滤显著降低了束硬化误差,从 9.5%降至 1.4%。相应的低千伏管负载(112mAs)是参考情况(21mAs)的五倍,但由于低千伏管负载与参考(100mAs)的高千伏管负载相当,因此它保持在一定的实用水平。束硬化减少的优越性反映在模拟图像中;例如,与参考过滤相比,通过使用 W 过滤器,皮质骨的模拟和理论电子密度值之间的束硬化引起的偏差从 7.4%降低到 1.2%,即使没有对束硬化进行校正。
在减少束硬化方面,低千伏 W 过滤的 DECT 更有效,可在管负载和剂量无额外增加的情况下实现电子密度的定量测量。