Pharm Res. 2012 Feb;29(2):345-51. doi: 10.1007/s11095-011-0559-6. Epub 2011 Aug 9.
To evaluate salicylic acid tablets as a candidate reference material in a Performance Verification Test (PVT) when a USP performance test for dissolution (General Chapter <711>) relies on USP Apparatus 4 (flow-through cell).
We developed a dissolution procedure relying on Apparatus 4 and salicylic acid tablets. Thereafter, a designed experiment was conducted to identify operational variables that significantly affect salicylic acid dissolution in this apparatus.
Four variables (size of glass beads, cell temperature, flow rate, level of deaeration) and one combination effect (deaeration/bead size) were significant for mean percent dissolved. Two variables (tablet orientation, level of deaeration) were significant for standard deviation results, but these effects were less pronounced than those for mean percent dissolved results. Three variables (analyst, tester manufacturer, amount of glass beads) had no statistically significant effects on either the mean or standard deviation of the responses.
The proposed PVT is capable of probing effects of changes in several critical operational parameters of Apparatus 4. Salicylic acid tablets were shown to be a suitable reference material for the PVT. The PVT using salicylic acid tablets satisfies important aspects of a PVT.
在依赖 USP 溶出度性能测试(一般章节 <711>)的 USP 性能测试(General Chapter <711>)中,评估水杨酸片作为性能验证测试(PVT)的候选参比物质,当该测试使用 USP 仪器 4(流通池)时。
我们开发了一种依赖于仪器 4 和水杨酸片的溶出度程序。此后,进行了一项设计实验,以确定对该仪器中水杨酸溶解有显著影响的操作变量。
四个变量(玻璃珠大小、池温度、流速、脱气程度)和一个组合效应(脱气/珠大小)对平均溶解百分比有显著影响。两个变量(片剂方向、脱气程度)对标准偏差结果有显著影响,但这些影响不如平均溶解百分比结果显著。三个变量(分析员、测试器制造商、玻璃珠的数量)对响应的平均值或标准偏差均无统计学显著影响。
所提出的 PVT 能够探测仪器 4 中几个关键操作参数变化的影响。水杨酸片被证明是 PVT 的合适参比物质。使用水杨酸片的 PVT 满足 PVT 的重要方面。