Dipartimento di Fisiologia e Farmacologia, V. Erspamer - Sapienza University of Rome and Centro Antidroga and Farmacologia Clinica Policlinico Umberto I, Italy.
J Pharm Biomed Anal. 2011 Dec 5;56(4):763-70. doi: 10.1016/j.jpba.2011.07.018. Epub 2011 Jul 23.
Methods based on HPLC technology are the most frequently adopted for monitoring blood levels of novel antiepileptics. Here a rapid method based on HPTLC was developed for quantitative determination of lamotrigine (LTG), zonisamide (ZNS) and levetiracetam (LVT) in human plasma and compared with HPLC and LC-MS/MS methods. Chromatographic separation was achieved on silical gel 60F(254) plates using ethylacetate:methanol:ammonia (91:10:15v/v/v) as mobile phase. Quantitative analysis was carried out by densitometry at a wavelength of 312, 240 and 210nm for LTG, ZNS and LVT, respectively. Calibration curves were linear over range of 0-200ng for LTG and ZNS and 0-400ng for and LVT. The limit of quantification of LTG, ZNS and LTV was found to be 3.69, 3.7 and 6.85μg/ml, respectively. Intra and inter-assay precision provided relative standard deviations lower than 10% for all three analytes. Correlation and Bland-Altman plot showed general agreement between HPTLC and LC-MS/MS quantification, with a mean bias of -0.25, -0.46 and 0.5μg/ml for LTG ZNS and LVT, respectively. Likewise, comparison between HPLC-UV and LC-MS/MS showed good agreement for all the three compounds analyzed. In conclusion, the proposed HPTLC method is simple, rapid, precise and accurate. It therefore is appropriate for the routine quantification of therapeutic levels of LTG, ZNS and LVT in human plasma.
方法基于高效液相色谱技术是最常用的监测新型抗癫痫药物的血药浓度。这里开发了一种快速的基于高效薄层色谱法的方法,用于定量测定拉莫三嗪(LTG)、佐米曲普坦(ZNS)和左乙拉西坦(LVT)在人血浆中,并与高效液相色谱法和液相色谱-串联质谱法(LC-MS/MS)进行了比较。色谱分离在硅胶 60F(254)板上进行,使用乙酸乙酯:甲醇:氨水(91:10:15v/v/v)作为流动相。定量分析在 312、240 和 210nm 处进行,分别用于 LTG、ZNS 和 LVT。LTG 和 ZNS 的校准曲线在 0-200ng 范围内线性,LVT 的校准曲线在 0-400ng 范围内线性。LTG、ZNS 和 LTV 的定量下限分别为 3.69、3.7 和 6.85μg/ml。对于所有三种分析物,内和日间精密度提供的相对标准偏差均低于 10%。相关性和 Bland-Altman 图表明 HPTLC 和 LC-MS/MS 定量之间具有总体一致性,对于 LTG、ZNS 和 LVT,平均偏差分别为-0.25、-0.46 和 0.5μg/ml。同样,HPLC-UV 和 LC-MS/MS 之间的比较表明,对于分析的所有三种化合物都具有良好的一致性。总之,所提出的 HPTLC 方法简单、快速、精确和准确。因此,适用于人血浆中 LTG、ZNS 和 LVT 的治疗水平的常规定量。