Regional Vascular Unit, Royal Liverpool University Hospital, Liverpool, UK.
J Endovasc Ther. 2011 Aug;18(4):569-75. doi: 10.1583/11-3471.1.
To quantify the compression force acting on target vessel stents as a consequence of the misalignment between the native aortic anatomy and the fenestrated stent-graft owing to measuring errors during the design of the device.
The material properties of a fenestrated Zenith stent-graft were determined using a standardized tensile testing protocol. Aortic anatomy was modeled using fresh porcine aortas that were subjected to tensile testing. The net force acting on a target vessel stent due to incremental discrepancy between the target vessel ostia and the stent-graft fenestrations was calculated as the difference in wall tension between the aorta and the stent-graft in diastole and systole. The change in diameter between diastole and systole was set to 8%.
Using the diastole model, underestimation of circumferential target vessel position by 15°, 22.5°, and 30° resulted in net forces on the target vessel stent of 0.6, 0.8, and 1.1 N, respectively. Overestimation of target vessel position by the same increments resulted in net forces of 0.3, 0.6, and 0.9 N, respectively. With the systolic model, underestimating target vessel position by 30° resulted in a 2.1-N maximum force on the stent, which potentially threatened the seal. In the longitudinal direction, underestimating target vessel separation by up to 10 mm resulted in a maximal force on the stent of 6.1 N, while overestimating target vessel separation did not result in any additional force on the stent due to fabric infolding.
The magnitude of the forces generated solely due to mismatch between stent-graft design and native anatomy is modest and is unlikely to cause significant deformation of target vessel stents. Mismatch, however, may cause loss of seal.
量化由于设备设计过程中的测量误差导致的原生主动脉解剖结构与开窗支架移植物之间的错位对目标血管支架所产生的压缩力。
采用标准化拉伸测试方案确定开窗型 Zenith 支架移植物的材料特性。使用经过拉伸测试的新鲜猪主动脉来模拟主动脉解剖结构。在舒张期和收缩期,计算由于目标血管口与支架移植物开窗之间的增量差异而作用于目标血管支架上的净力,其方法为主动脉和支架移植物之间的壁张力差。舒张期和收缩期之间的直径变化设定为 8%。
使用舒张期模型,当目标血管位置在圆周方向上被低估 15°、22.5°和 30°时,目标血管支架上的净力分别为 0.6、0.8 和 1.1 N。当目标血管位置被高估相同增量时,目标血管支架上的净力分别为 0.3、0.6 和 0.9 N。在收缩期模型中,当目标血管位置被低估 30°时,支架上会产生 2.1 N 的最大力,这可能会威胁到密封效果。在纵向方向上,当目标血管分离被低估最多 10 mm 时,支架上会产生 6.1 N 的最大力,而高估目标血管分离不会导致支架上产生任何额外的力,这是因为织物内折。
仅仅由于支架移植物设计与原生解剖结构不匹配而产生的力的大小适中,不太可能导致目标血管支架发生显著变形。然而,不匹配可能会导致密封失效。