Intraluminal vascular ultrasound: preliminary report of dimensional and morphologic accuracy.

作者信息

Kopchok G E, White R A, Guthrie C, Hsiang Y, Rosenbaum D, White G H

机构信息

Department of Surgery, Harbor-UCLA Medical Center, Torrance 90509.

出版信息

Ann Vasc Surg. 1990 May;4(3):291-6. doi: 10.1007/BF02009460.

Abstract

The role of intraluminal ultrasound for diagnosis and monitoring treatment of vascular disease has yet to be defined. This study evaluated the dimensional precision and morphologic accuracy of an intraluminal ultrasound system which consists of a 5.5 French external diameter ultrasonic catheter with a central lumen for passage of a guidewire. Ultrasound images from five in-vitro human and three porcine arterial segments and two in-vivo arteriosclerotic canine arteries were compared to dimensions obtained from arteriograms and from sections of the specimens. Each gross and histological specimen and ultrasound image was scaled, photographed and enlarged up to 20 times and measured for vessel intraluminal and adventitial or outer diameter and wall thickness. Intraluminal and outer diameters and wall thickness from normal in-vitro specimens correlated significantly with dimensions obtained from histologic specimens (r = 0.99, p less than 0.005 for internal and outer diameters and r = 0.73, p less than 0.005 for wall thickness). The mean differences of luminal diameters measured from the vessels supported within a silicone rubber mold was 0.05 +/- 0.09 mm (n = 20). Diameters of outer diameter and wall thickness were less reliably defined, the average margin of error being 0.49 +/- 0.39 mm and 0.29 +/- 0.26 mm, respectively. The mean difference between in-vivo ultrasound and arteriographic diameters was 0.61 +/- 0.38 mm (n = 12). Correlation of luminal diameters between ultrasound and arteriogram was significant (r = 0.76, p less than 0.02). The ultrasound images also differentiated a laminated appearance of normal vessel anatomy from non-uniform or dense signals seen in atherosclerotic lesions.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

腔内超声在血管疾病诊断和治疗监测中的作用尚未明确。本研究评估了一种腔内超声系统的尺寸精度和形态学准确性,该系统由一根外径为5.5法国规格的超声导管组成,导管有一个中心腔用于导丝通过。将来自五个体外人体和三个猪动脉段以及两条体内动脉粥样硬化犬动脉的超声图像与动脉造影和标本切片获得的尺寸进行比较。对每个大体标本、组织学标本和超声图像进行缩放、拍照并放大20倍,测量血管腔内、外膜或外径以及壁厚。体外正常标本的内径、外径和壁厚与组织学标本获得的尺寸显著相关(内径和外径的r = 0.99,p < 0.005;壁厚的r = 0.73,p < 0.005)。在硅橡胶模具中支撑的血管所测量的管腔直径平均差异为0.05±0.09毫米(n = 20)。外径和壁厚的定义不太可靠,平均误差分别为0.49±0.39毫米和0.29±0.26毫米。体内超声与动脉造影直径之间的平均差异为0.61±0.38毫米(n = 12)。超声与动脉造影之间的管腔直径相关性显著(r = 0.76,p < 0.02)。超声图像还能区分正常血管解剖结构的分层外观与动脉粥样硬化病变中所见的不均匀或密集信号。(摘要截断于250字)

相似文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验