Institute of Radiology, University Medical Center Regensburg, Regensburg, Germany.
Clin Hemorheol Microcirc. 2011;48(1):187-98. doi: 10.3233/CH-2011-1399.
Postoperative monitoring of transplanted free flaps is an essential tool to reveal possible complications. The aim of this study was to compare the value of time-intensity-curve- (TIC-) analysis based on grey scale data of contrast-enhanced ultrasound (CEUS) and dynamic contrast-enhanced (DCE-) magnetic resonance imaging (MRI). Postoperative imaging was performed in 11 patients following free flap transplantation and TIC analysis was obtained in identical regions of interest (ROI) of CEUS and DCE MRI data. Microcirculation was assessed in superficial (0-1 cm), middle (1-2 cm), and deep (2-3 cm) ROIs in one or two different positions within the flap resulting in a total of 46 ROIs evaluated (in very thin flaps only superficial and middle ROIs were assessed). For both imaging methods, mean signal increase was found to be significantly higher in ROIs of normally perfused flaps (n = 40) compared to ROIs with compromised microcirculation (n = 6). Although TIC analysis allows quantification of microcirculation in different regions of the flap, in this preliminary study no distinct threshold could be defined to differentiate flaps with normal and compromised microcirculation.
术后游离皮瓣监测是揭示可能并发症的重要手段。本研究旨在比较基于灰阶数据的超声造影(CEUS)和动态对比增强磁共振成像(DCE MRI)的时间-强度曲线(TIC)分析的价值。11 例游离皮瓣移植术后进行了术后影像学检查,并在 CEUS 和 DCE MRI 数据的相同感兴趣区(ROI)中获得了 TIC 分析。在皮瓣的一个或两个不同位置评估了浅层(0-1cm)、中层(1-2cm)和深层(2-3cm)ROI 的微循环,总共评估了 46 个 ROI(非常薄的皮瓣仅评估浅层和中层 ROI)。对于两种成像方法,与微循环受损的 ROI(n=6)相比,正常灌注皮瓣的 ROI(n=40)的平均信号增加明显更高。尽管 TIC 分析允许定量评估皮瓣不同区域的微循环,但在这项初步研究中,无法定义明确的阈值来区分正常和微循环受损的皮瓣。