使用氨选择性电极对血液或血浆的高氯酸上清液中的氨进行连续流动分析。

Continuous-flow analysis of ammonia in perchloric acid supernate of blood or plasma using an ammonia-selective electrode.

作者信息

Moses G C, Thibert R J, Draisey T F

出版信息

J Clin Pathol. 1978 Dec;31(12):1207-11. doi: 10.1136/jcp.31.12.1207.

Abstract

A simple automated method for the estimation of ammonia in perchloric acid supernate of blood or plasma using an ion-selective electrode (Orion Ammonia-selective electrode, Model 95-10) is described. The reliability of the proposed method has been checked against an ion-exchange resin procedure, which has been chosen as a standard procedure. Regression equation and correlation coefficient for the proposed method are y = 0.7 X + 10 and 0.945, respectively, as compared with the chosen standard method. Within-run and between-run precision are 2.1% and 3.5% respectively. The average percent recovery is 97.5% and a tentative range is 13-73 microgram/dl (9-52 micronmol/l) ammonia nitrogen.

摘要

描述了一种使用离子选择电极(奥立龙氨选择性电极,型号95 - 10)测定血液或血浆高氯酸上清液中氨的简单自动化方法。已将所提出方法的可靠性与作为标准程序选择的离子交换树脂程序进行了核对。与所选标准方法相比,所提出方法的回归方程和相关系数分别为y = 0.7X + 10和0.945。批内精密度和批间精密度分别为2.1%和3.5%。平均回收率为97.5%,暂定范围为13 - 73微克/分升(9 - 52微摩尔/升)氨氮。

相似文献

7
Evaluation of the enzymatic ammonia method for urine on the Du Pont automatic clinical analyzer.
J Clin Chem Clin Biochem. 1978 May;16(5):289-91. doi: 10.1515/cclm.1978.16.5.289.
8
An automated microassay for blood ammonia.一种用于血氨的自动化微量测定法。
J Clin Chem Clin Biochem. 1979 Apr;17(4):247-50. doi: 10.1515/cclm.1979.17.4.247.
10
[The hyperammonemia of exercise; value of automatic measurement].[运动性高氨血症;自动测量的价值]
Arch Int Physiol Biochim Biophys. 1993 Sep-Oct;101(5):311-4. doi: 10.3109/13813459309003930.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验