Vieli A
Institute of Biomedical Engineering and Medical Informatics, University of Zürich, Switzerland.
Urol Int. 1990;45(4):251-7. doi: 10.1159/000281717.
In order to utilize Doppler modalities in an optimal manner, one should both be aware of their potential and respect their limitations. To this end, a variety of technical and physical factors affecting the performance of Doppler devices are discussed: (1) the velocity of propagation limiting the temporal resolution of pulsed systems; (2) intervening tissue that may weaken the echo considerably; (3) the size of the sample volume that could be too small to provide a transvascular mean velocity or too large to discriminate flow in adjacent vessels; (4) the principles of continuous wave mode lacking depth resolution, and pulsed mode being prone to aliasing; (5) the additional complexity of duplex/triplex systems; (6) the tight temporal restrictions of 2-dimensional color mappers which cause poor velocity resolution, and (7) the angle of incidence that needs to be assessed in order to measure absolute velocity and the cross-sectional area that must be measured additionally for the calculation of the volume flow rate. Based on this discussion, some hints on selecting instruments or instrument parameters are given.
为了以最佳方式利用多普勒技术,人们既应了解其潜力,也应尊重其局限性。为此,讨论了影响多普勒设备性能的各种技术和物理因素:(1)传播速度限制了脉冲系统的时间分辨率;(2)中间组织可能会使回声显著减弱;(3)样本容积的大小可能过小而无法提供跨血管平均速度,或过大而无法区分相邻血管中的血流;(4)连续波模式缺乏深度分辨率,脉冲模式容易出现混叠;(5)双功/三功系统的额外复杂性;(6)二维彩色成像仪严格的时间限制导致速度分辨率较差,以及(7)为了测量绝对速度需要评估入射角,为了计算体积流量还必须额外测量横截面积。基于此讨论,给出了一些选择仪器或仪器参数的提示。