A modified device for intraoral radiography to assess the distal osseous defects of mandibular second molar after impacted third molar surgery.

作者信息

Ana Faria-Inocencio, Mercedes Gallas-Torreira

机构信息

Stomatology Department, Faculty of Medicine and Dentistry, Santiago de Compostela University, Santiago de Compostela, Spain.

出版信息

Imaging Sci Dent. 2011 Sep;41(3):115-21. doi: 10.5624/isd.2011.41.3.115. Epub 2011 Sep 15.

Abstract

PURPOSE

This article is to describe a modified device for intraoral radiography which was developed to obtain reproducible radiographic images for assessment of distal osseous defects of the mandibular second molar (2 Mm) after impacted third molar (3 Mm) surgery.

MATERIALS AND METHODS

A commercial available alignment system for posterior region was modified by adding a reference gauge pin (millimetric) and threading a hollow acrylic cylinder at the ring of the radiographic positioner to attach the X-ray collimator. The design included customized resin acrylic stent for the occlusal surface of the 2Mm in maximum intercuspal position, individualizing the biteblock positioner. Periapical radiographs were taken before and after surgical extraction of 3 Mm, employing the radiographic technique of parallelism described by Kugelberg (1986) with this modified film holder and inserting the gauge pin on the deepest bone probing depth point.

RESULTS

This technique permitted to obtain standardized periapical radiographs with a moderate to high resolution, repeatability, and accuracy. There was no difference between the measurements on the pre- and post-operative radiographs. This technique allowed better maintenance of the same geometric position compared with conventional one. The insertion of the gauge pin provided the same reference point and localized the deepest osseous defect on the two-dimensional radiographs.

CONCLUSION

This technique allowed better reproducibility in posterior radiographic records (distal surface of 2 Mm) and more accurate measurements of radiographic bone level by the use of a millimetric pin.

摘要

目的

本文旨在描述一种改良的口腔内放射摄影设备,该设备旨在获取可重复的放射影像,用于评估阻生第三磨牙(3Mm)手术后下颌第二磨牙(2Mm)远中骨缺损情况。

材料与方法

对一种市售的后部区域定位系统进行改良,增加一个参考量规针(毫米级),并在放射摄影定位器的环上穿一个中空的丙烯酸圆柱体以连接X射线准直器。该设计包括为处于最大牙尖交错位的2Mm咬合面定制的树脂丙烯酸支架,使咬块定位器个性化。在3Mm手术拔除前后拍摄根尖片,采用Kugelberg(1986)描述的平行放射摄影技术,使用这种改良的胶片夹,并将量规针插入最深骨探测深度点。

结果

该技术能够获得具有中等到高分辨率、可重复性和准确性的标准化根尖片。术前和术后放射片上的测量结果没有差异。与传统技术相比,该技术能更好地保持相同的几何位置。量规针的插入提供了相同的参考点,并在二维放射片上定位了最深的骨缺损。

结论

该技术通过使用毫米级针,能在后部放射记录(2Mm远中面)中实现更好的可重复性,并能更准确地测量放射学骨水平。