Suppr超能文献

因同轴(贝恩)回路故障导致的高碳酸血症。

Hypercapnia resulting from a faulty co-axial (Bain) circuit.

作者信息

Singh Ishwar, Gupta Monika, Singh Tarun K

机构信息

Department of Anaesthesia, Jaipur Golden Hospital, New Delhi, India.

出版信息

Indian J Anaesth. 2011 Jul;55(4):402-4. doi: 10.4103/0019-5049.84853.

Abstract

The Bain co-axial circuit is fully established in general anaesthesia practice. A major concern is the potential malfunctioning of the circuit due to avulsion of the inner fresh gas delivery tube at the machine end of the circuit. The following case report presents a case in which a patient connected to the Bain circuit developed severe hypercapnia in the early intraoperative period due to the above mentioned defect.

摘要

贝恩同轴回路在全身麻醉实践中已被广泛应用。一个主要问题是,回路在麻醉机端的内新鲜气体输送管可能因撕裂而导致回路出现潜在故障。以下病例报告介绍了一例连接到贝恩回路的患者,在术中早期因上述缺陷出现严重高碳酸血症的情况。