Tsroya S, Pelled O, German U, Katorza E, Abraham A, Alfassi Z B
Nuclear Research Center Negev, Beer Sheva, Israel.
Appl Radiat Isot. 2012 Feb;70(2):397-403. doi: 10.1016/j.apradiso.2011.10.002. Epub 2011 Oct 19.
The color quenching correction for Cerenkov counting is normally based on indexes defined for liquid scintillation spectra. Because of the differences in the spectrum shapes, these indexes are not optimal, and a different method was proposed based on the integral counting of the external (152)Eu source spectra of a 1200 Quantulus system. We have shown that this method, named ESAR (external source area ratio), is superior to all the other correction methods, offering the highest sensitivity and the widest range of application.
切伦科夫计数的颜色淬灭校正通常基于为液体闪烁光谱定义的指标。由于光谱形状的差异,这些指标并非最优,因此提出了一种基于1200 Quantulus系统外部(152)Eu源光谱积分计数的不同方法。我们已经表明,这种名为ESAR(外源源面积比)的方法优于所有其他校正方法,具有最高的灵敏度和最广泛的应用范围。