肾小球滤过率预测方程在肾移植中的准确性:Architect c8000 上新的胶乳增强散射比浊法测定胱抑素 C 的验证。

Accuracy of GFR predictive equations in renal transplantation: validation of a new turbidimetric cystatin C assay on Architect c8000.

机构信息

Department of Biochemistry, University of Montpellier 1, Montpellier, France.

出版信息

Clin Biochem. 2012 Jan;45(1-2):151-3. doi: 10.1016/j.clinbiochem.2011.10.018. Epub 2011 Nov 6.

Abstract

OBJECTIVES

To evaluate the Sentinel-PETIA cystatin C on Architect c8000 analyzer.

DESIGN AND METHODS

We assessed analytical performances and clinical relevance by comparison with a reference isotopic method in kidney transplant recipients.

RESULTS

This assay exhibited reliable precision and was close to the non standardized Siemens-PENIA method. All tested equations allowed reliable assessment of GFR.

CONCLUSIONS

Cystatin C improved GFR determination at the critical level of 60 mL/min/1.73 m². New formulas might be necessary after IFCC standardization.

摘要

目的

评估 Architect c8000 分析仪上的 Sentinel-PETIA 胱抑素 C。

设计和方法

我们通过与肾移植受者的参考同位素方法进行比较,评估了分析性能和临床相关性。

结果

该测定法具有可靠的精密度,且接近非标准化的西门子-PENIA 方法。所有测试的方程都允许可靠地评估 GFR。

结论

胱抑素 C 在 60 mL/min/1.73 m² 的关键水平提高了 GFR 测定的准确性。在 IFCC 标准化后,可能需要新的公式。