核医学成像多中心试验标准化建议:建立欧洲 123I-FP-CIT SPECT 数据库的前提条件。

Proposal for the standardisation of multi-centre trials in nuclear medicine imaging: prerequisites for a European 123I-FP-CIT SPECT database.

机构信息

Institute of Nuclear Medicine, UCLH NHS Foundation Trust and University College London, London, UK.

出版信息

Eur J Nucl Med Mol Imaging. 2012 Jan;39(1):188-97. doi: 10.1007/s00259-011-1884-z.

Abstract

PURPOSE

Multi-centre trials are an important part of proving the efficacy of procedures, drugs and interventions. Imaging components in such trials are becoming increasingly common; however, without sufficient control measures the usefulness of these data can be compromised. This paper describes a framework for performing high-quality multi-centre trials with single photon emission computed tomography (SPECT), using a pan-European initiative to acquire a normal control dopamine transporter brain scan database as an example.

METHODS

A framework to produce high-quality and consistent SPECT imaging data was based on three key areas: quality assurance, the imaging protocol and system characterisation. Quality assurance was important to ensure that the quality of the equipment and local techniques was good and consistently high; system characterisation helped understand and where possible match the performance of the systems involved, whereas the imaging protocol was designed to allow a degree of flexibility to best match the characteristics of each imaging device.

RESULTS

A total of 24 cameras on 15 sites from 8 different manufacturers were evaluated for inclusion in our multi-centre initiative. All results matched the required level of specification and each had their performance characterised. Differences in performance were found between different system types and cameras of the same type. Imaging protocols for each site were modified to match their individual characteristics to produce comparable high-quality SPECT images.

CONCLUSION

A framework has been designed to produce high-quality data for multi-centre SPECT studies. This framework has been successfully applied to a pan-European initiative to acquire a healthy control dopamine transporter image database.

摘要

目的

多中心试验是证明程序、药物和干预措施疗效的重要组成部分。此类试验中的成像组件越来越常见;然而,如果没有足够的控制措施,这些数据的有用性可能会受到影响。本文以一个泛欧倡议为例,描述了一个用于进行高质量单光子发射计算机断层扫描(SPECT)多中心试验的框架。

方法

一个用于生成高质量和一致的 SPECT 成像数据的框架基于三个关键领域:质量保证、成像协议和系统特性。质量保证对于确保设备和本地技术的质量良好且始终保持高水平非常重要;系统特性有助于了解并在可能的情况下匹配所涉及系统的性能,而成像协议旨在允许一定程度的灵活性,以最佳匹配每个成像设备的特性。

结果

共有来自 8 家制造商的 15 个地点的 24 台相机被评估是否纳入我们的多中心计划。所有结果均符合所需的规格要求,并且每个结果都进行了性能特征描述。不同系统类型和同一类型的相机之间存在性能差异。为每个站点修改了成像协议,以匹配其各自的特性,从而生成可比较的高质量 SPECT 图像。

结论

设计了一个框架来生成高质量的数据用于多中心 SPECT 研究。该框架已成功应用于一个泛欧倡议,以获取健康对照多巴胺转运蛋白图像数据库。