Bardy G H, Allen M D, Mehra R, Johnson G
Department of Medicine, University of Washington, Seattle.
J Am Coll Cardiol. 1990 Oct;16(4):887-95. doi: 10.1016/s0735-1097(10)80337-7.
With use of a coronary sinus catheter electrode, a right ventricular catheter electrode and a chest wall patch electrode system, defibrillation threshold voltage, current and energy were measured with four distinct transvenous defibrillation techniques delivered in random sequence in each of 12 survivors of cardiac arrest immediately before implantation of a standard epicardial patch defibrillation system. The four transvenous defibrillation techniques were 1) single pathway monophasic pulsing, 2) single pathway biphasic pulsing, 3) dual pathway sequential pulsing, and 4) dual pathway simultaneous pulsing. A transvenous defibrillation method was considered to be potentially useful only if the defibrillation threshold was less than or equal to 500 V (less than or equal to 15 J delivered energy). The 500 V value would allow a 2:1 defibrillation safety margin for a device with a maximal output of 30 J. No single transvenous pulsing technique was uniformly superior in efficacy. However, by choosing the best pulsing technique for each patient, it was possible to obtain an average defibrillation threshold of 410 +/- 135 V leading edge voltage, 7.2 +/- 2.5 A leading edge current and 11.3 +/- 7.4 J delivered energy for the group of 12 patients. With the ability to vary defibrillation technique, transvenous antiarrhythmic device implantation would have been possible in 10 (83%) of the 12 patients at or below a 15 J defibrillation threshold cutoff point. In contrast, if only one transvenous defibrillation method had been used, as few as 5 and at most 8 of the 12 patients would have been candidates for a transvenous defibrillation system given a 15 J defibrillation threshold cutoff point for insertion.(ABSTRACT TRUNCATED AT 250 WORDS)
在12名心脏骤停幸存者植入标准心外膜贴片除颤系统之前,使用冠状窦导管电极、右心室导管电极和胸壁贴片电极系统,以随机顺序采用四种不同的经静脉除颤技术测量除颤阈值电压、电流和能量。这四种经静脉除颤技术分别为:1)单路径单相脉冲;2)单路径双相脉冲;3)双路径顺序脉冲;4)双路径同步脉冲。仅当除颤阈值小于或等于500V(输送能量小于或等于15J)时,经静脉除颤方法才被认为可能有用。500V的值将为最大输出为30J的设备提供2:1的除颤安全裕度。没有一种经静脉脉冲技术在疗效上始终具有优势。然而,通过为每位患者选择最佳脉冲技术,12名患者组的平均除颤阈值可达到前沿电压410±135V、前沿电流7.2±2.5A和输送能量11.3±7.4J。由于能够改变除颤技术,12名患者中有10名(83%)在15J除颤阈值截止点或以下时有可能植入经静脉抗心律失常装置。相比之下,如果仅使用一种经静脉除颤方法,对于15J除颤阈值截止点的植入而言,12名患者中只有5至8名可能成为经静脉除颤系统的候选者。(摘要截断于250字)