Prospective comparison of sequential pulse and single pulse defibrillation with use of two different clinically available systems.

作者信息

Bardy G H, Ivey T D, Allen M D, Johnson G, Greene H L

机构信息

Department of Medicine, University of Washington, Seattle.

出版信息

J Am Coll Cardiol. 1989 Jul;14(1):165-71. doi: 10.1016/0735-1097(89)90068-5.

Abstract

Sixteen out-of-hospital survivors of ventricular fibrillation underwent a prospective, randomized, intraoperative comparison of sequential pulse and single pulse defibrillation with use of two distinct electrode systems and waveform shapes currently available for clinical use. Defibrillation was tested alternately with either the single pulse or the sequential pulse system 10 s into an episode of ventricular fibrillation. Sequential pulse defibrillation was performed with two 4 ms truncated exponential pulses of constant duration delivered to three equally spaced oval epicardial patch electrodes composed of concentric coils. The posterior left ventricular electrode served as the common cathode. The first anode was over the anterior right ventricle and the second anode was over the anterior left ventricle. Single pulse defibrillation was performed with the standard intracardiac defibrillation system with use of a single truncated exponential pulse with a fixed 65% tilt delivered across two rectangular, wire mesh epicardial patch electrodes positioned over the anterior right ventricle and posterolateral left ventricle. During defibrillation threshold determination, voltage and current waveforms were recorded and used to determine pulsing resistance and delivered and stored energy. Average defibrillation threshold leading edge voltage for the single pulse technique was 273 +/- 101 V compared with 246 +/- 67 V (11% less) for the sequential pulse technique (p = 0.136). Defibrillation threshold leading edge current for the single pulse technique was 6.7 +/- 2.5 A compared with 5.2 +/- 1.7 A (29% less) for the sequential pulse method (p = 0.005). The defibrillation threshold delivered energy was 5.6 +/- 4.0 J for the single pulse technique and 3.5 +/- 1.8 J (38% less) for the sequential pulse technique (p = 0.021).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验