Abstract

INTRODUCTION

A number of rotary nickel-titanium (NiTi) systems have been developed to provide better, faster, and easier cleaning and shaping of the root canal system, and recently, rotary NiTi systems designed for root canal retreatment have been introduced. Because the main problem with the rotary NiTi files is fracture, the aim of this study was to compare the cyclic fatigue resistance of 3 different rotary NiTi systems designed for root canal retreatment.

METHODS

Total of 60 instruments of 3 different rotary NiTi systems designed for root canal retreatment were used in this study. Twenty R-Endo R3, 20 ProTaper D3, and 20 Mtwo R (Retreatment) 25.05 instruments were tested. Cyclic fatigue testing of instruments was performed by using a device that allowed the instruments to rotate freely inside an artificial canal. Each instrument was rotated until fracture occurred, and the number of cycles to fracture for each instrument was calculated. Representative samples were also evaluated under a scanning electron microscope to confirm the fracture was flexural. Data were analyzed by using 1-way analysis of variance test.

RESULTS

R-Endo R3 instruments showed better cyclic fatigue resistance than ProTaper D3 and Mtwo R 25.05 instruments, and the difference was statistically significant (P < .05). There was no significant difference between ProTaper D3 and Mtwo R 25.05 groups.

CONCLUSIONS

The R-Endo R3 instruments were more resistant to fatigue failure than ProTaper D3 and Mtwo R 25.05.

摘要

简介

为了更好、更快、更容易地清理和成形根管系统,已经开发了许多旋转镍钛(NiTi)系统,最近,专门用于根管再治疗的旋转 NiTi 系统也已推出。由于旋转 NiTi 锉的主要问题是断裂,因此本研究的目的是比较 3 种不同的专门用于根管再治疗的旋转 NiTi 系统的抗循环疲劳性能。

方法

本研究共使用了 60 根 3 种不同的专门用于根管再治疗的旋转 NiTi 系统的器械。共测试了 20 根 R-Endo R3、20 根 ProTaper D3 和 20 根 Mtwo R(再治疗)25.05 器械。通过允许器械在人工根管内自由旋转的设备对器械进行循环疲劳测试。每个器械旋转至断裂,并计算每个器械的断裂循环次数。还使用扫描电子显微镜对代表性样本进行评估,以确认断裂是弯曲的。使用单向方差分析检验对数据进行分析。

结果

R-Endo R3 器械的抗循环疲劳性能优于 ProTaper D3 和 Mtwo R 25.05 器械,差异具有统计学意义(P <.05)。ProTaper D3 和 Mtwo R 25.05 组之间无显著差异。

结论

R-Endo R3 器械比 ProTaper D3 和 Mtwo R 25.05 器械更能抵抗疲劳失效。