School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham, B15 2TT, UK.
J Biomech. 2012 Feb 2;45(3):504-9. doi: 10.1016/j.jbiomech.2011.11.045. Epub 2011 Dec 15.
Total disc arthroplasty (TDA) can be used to replace a degenerated intervertebral disc in the spine. There are different designs of prosthetic discs, but one of the most common is a ball-and-socket combination. Contact between the bearing surfaces can result in high frictional torque, which can then result in wear and implant loosening. This study was designed to determine the effects of ball radius on friction. Generic models of metal-on-metal TDA were manufactured with ball radii of 10, 12, 14 and 16 mm, with a radial clearance of 0.015 mm. A simulator was used to test each sample in flexion-extension, lateral bending and axial rotation at frequencies of 0.25, 0.5, 0.75, 1, 1.25, 1.5, 1.75 and 2 Hz under loads of 50, 600, 1200 and 2000 N, in new born calf serum. Frictional torque was measured and Stribeck curves were plotted to illustrate the lubrication regime in each case. It was observed that implants with a smaller ball radius showed lower friction and showed boundary and mixed lubrication regimes, whereas implants with larger ball radius showed boundary lubrication only. This study suggests designing metal-on-metal TDAs with ball radius of 10 or 12 mm, in order to reduce wear and implant loosening.
全椎间盘置换术(TDA)可用于置换脊柱中退变的椎间盘。有不同设计的假体椎间盘,但其中最常见的是球窝组合。承载面之间的接触会导致高摩擦扭矩,从而导致磨损和植入物松动。本研究旨在确定球半径对摩擦的影响。制造了具有 10、12、14 和 16mm 球半径的通用金属对金属 TDA 模型,径向间隙为 0.015mm。模拟器用于在 50、600、1200 和 2000N 载荷下,以 0.25、0.5、0.75、1、1.25、1.5、1.75 和 2Hz 的频率,在屈伸、侧屈和轴向旋转下测试每个样本,在新生小牛血清中。测量了摩擦扭矩,并绘制了 Stribeck 曲线以说明每种情况下的润滑状态。观察到,球半径较小的植入物显示出较低的摩擦力,并显示出边界和混合润滑状态,而球半径较大的植入物仅显示边界润滑。本研究建议设计具有 10 或 12mm 球半径的金属对金属 TDA,以减少磨损和植入物松动。