To evaluate corneal hysteresis (CH) and corneal resistance factor (CRF) in keratoconic (KC) eyes before and after corneal collagen cross-linking (CXL). Furthermore, to determine potential correlations with a series of corneal and demographic factors.
The study consisted of 50 KC eyes that underwent CXL. CH and CRF were measured by the ocular response analyzer (ORA). Correlations were attempted with uncorrected visual acuity (UVA), best spectacle-corrected visual acuity (BSCVA), central corneal thickness (CCT), mean keratometry (Km), astigmatism (Astig.), residual astigmatism, age, and gender. Fifty non-KC eyes served as controls.
CH and CRF (mean ± SD) for non-KC eyes were 10.1 ± 1.9 mmHg and 9.7 ± 2.4 mmHg respectively, while for KC eyes preoperatively they were 8.2 ± 1.4 mmHg (p = 0.007) and 7.4 ± 2.3 mmHg (p = 0.01) respectively. Non-significant differences were detected between preoperative and postoperative CH and CRF measurements in KC eyes (p = 0.518 and p = 0.479 respectively). Significant correlations were found between ORA parameters and BSCVA, CCT, Km, Astig. and residual astigmatism.
ORA parameters demonstrate significant differences between KC and non-KC eyes. Both CH and CRF present significant correlations with visual acuity and corneal parameters. CXL exerts a non-significant impact on ORA measurements.
评估角膜滞后(CH)和角膜阻力因子(CRF)在圆锥角膜(KC)眼行角膜交联(CXL)前后的变化,并确定与一系列角膜和人口统计学因素的潜在相关性。
本研究共纳入 50 例接受 CXL 的 KC 眼。使用眼反应分析仪(ORA)测量 CH 和 CRF。尝试与未矫正视力(UVA)、最佳矫正视力(BSCVA)、中央角膜厚度(CCT)、平均角膜曲率(Km)、散光(Astig.)、残余散光、年龄和性别进行相关性分析。50 例非 KC 眼作为对照。
非 KC 眼的 CH 和 CRF(平均值±标准差)分别为 10.1±1.9mmHg 和 9.7±2.4mmHg,而 KC 眼术前分别为 8.2±1.4mmHg(p=0.007)和 7.4±2.3mmHg(p=0.01)。KC 眼术前和术后的 CH 和 CRF 测量值无显著差异(p=0.518 和 p=0.479 分别)。ORA 参数与 BSCVA、CCT、Km、Astig.和残余散光之间存在显著相关性。
ORA 参数在 KC 和非 KC 眼之间存在显著差异。CH 和 CRF 与视力和角膜参数均有显著相关性。CXL 对 ORA 测量值无显著影响。