Abbot Vascular, Santa Clara, CA, USA.
EuroIntervention. 2012 Feb;7(10):1227-35. doi: 10.4244/EIJV7I10A195.
Optical coherence tomography (OCT) of a bioresorbable vascular scaffold (BVS) produces a highly reflective signal outlining struts. This signal interferes with the measurement of strut thickness, as the boundaries cannot be accurately identified, and with the assessment of coverage, because the neointimal backscattering convolutes that of the polymer, frequently making them indistinguishable from one another. We hypothesise that Gaussian line spread functions (LSFs) can facilitate identification of strut boundaries, improving the accuracy of strut thickness measurements and coverage assessment.
Forty-eight randomly selected BVS struts from 12 patients in the ABSORB Cohort B clinical study and four Yucatan minipigs were analysed at baseline and follow-up (six months in humans, 28 days in pigs). Signal intensities from the raw OCT backscattering were fit to Gaussian LSFs for each interface, from which peak intensity and full-width-at-half-maximum (FWHM) were calculated. Neointimal coverage resulted in significantly different LSFs and higher FWHM values relative to uncovered struts at baseline (p<0.0001). Abluminal polymer-tissue interfaces were also significantly different between baseline and follow-up (p=0.0004 in humans, p<0.0001 in pigs). Using the location of the half-max of the LSF as the polymer-tissue boundary, the average strut thickness was 158±11 µm at baseline and 152±20 µm at six months (p=0.886), not significantly different from nominal strut thickness.
Fitting the raw OCT backscattering signal to a Gaussian LSF facilitates identification of the interfaces between BVS polymer and lumen or tissue. Such analysis enables more precise measurement of the strut thickness and an objective assessment of coverage.
生物可吸收血管支架(BVS)的光相干断层扫描(OCT)产生一个高度反射信号,勾勒出支架的轮廓。这个信号会干扰支架厚度的测量,因为边界无法准确识别,也会干扰覆盖率的评估,因为新生内膜的背向散射会使聚合物的背向散射变得模糊,经常使它们难以区分。我们假设高斯线扩展函数(LSF)可以帮助识别支架的边界,从而提高支架厚度测量和覆盖率评估的准确性。
从 ABSORB 队列 B 临床试验中的 12 名患者的 48 个随机选择的 BVS 支架和 4 只尤卡坦迷你猪在基线和随访时(人类为 6 个月,猪为 28 天)进行了分析。对原始 OCT 背向散射的信号强度进行了高斯 LSF 拟合,计算了每个界面的峰值强度和半峰全宽(FWHM)。新生内膜覆盖率导致基线时未覆盖的支架的 LSF 和更高的 FWHM 值有显著差异(p<0.0001)。在基线时,外膜聚合物-组织界面也与随访时显著不同(p=0.0004 人,p<0.0001 猪)。使用 LSF 半最大值的位置作为聚合物-组织边界,支架的平均厚度在基线时为 158±11 µm,在 6 个月时为 152±20 µm(p=0.886),与标称支架厚度无显著差异。
将原始 OCT 背向散射信号拟合到高斯 LSF 中,有助于识别 BVS 聚合物与管腔或组织之间的界面。这种分析可以更精确地测量支架的厚度,并对覆盖率进行客观评估。