Routine chest and abdominal high-pitch CT: an alternative low dose protocol with preserved image quality.

机构信息

Institute of Diagnostic and Interventional Radiology, University Hospital Zürich, Rämistrasse 101, Zürich, CH-8091, Switzerland.

出版信息

Eur J Radiol. 2012 Mar;81(3):e392-7. doi: 10.1016/j.ejrad.2011.12.017. Epub 2012 Jan 9.

Abstract

OBJECTIVE

To investigate the radiation dose and image quality of the high-pitch dual source computer tomography (DSCT) for routine chest and abdominal scans.

METHODS

130 consecutive patients (62 female, 68 male, median age 55 years) were included. All patients underwent 128-slice high-pitch DSCT (chest n=99; abdomen n=84) at a pitch of 3.2. Two observers independently rated image quality using a 4-point score (1: excellent to 4: non-diagnostic). Image noise was measured and operational radiation dose quantities were recorded. An additional group of 132 patients (chest, n=80; abdomen n=52) scanned with standard-pitch CT matched for age, gender, and body mass index (BMI) served as control group.

RESULTS

Interobserver agreement for image quality rating was good (k=0.74). Subjective image quality of high-pitch CT was diagnostic in all patients (median score chest; 2, median score abdomen: 2). Image noise of high-pitch CT was comparable to standard-pitch for the chest (p=0.32) but increased in the abdomen (p<0.0001). For high-pitch CT radiation dose was 4.4±0.9 mSv (chest) and 6.5±1.2 mSv (abdomen). These values were significantly lower compared to standard-pitch CT (chest: 5.5±1.2 mSv; abdomen: 11.3±3.8 mSv).

CONCLUSION

Based on the technical background high-pitch dual source CT may serve as an alternative scan mode for low radiation dose routine chest and abdominal CT.

摘要

目的

研究高螺距双源计算机断层扫描(DSCT)在常规胸部和腹部扫描中的辐射剂量和图像质量。

方法

纳入 130 例连续患者(62 例女性,68 例男性,中位年龄 55 岁)。所有患者均接受 128 层高螺距 DSCT(胸部 n=99;腹部 n=84)扫描,螺距为 3.2。两名观察者独立使用 4 分制(1:优秀至 4:无法诊断)对图像质量进行评分。测量图像噪声并记录操作辐射剂量。另外一组 132 例年龄、性别和体重指数(BMI)匹配的常规螺距 CT 扫描患者(胸部 n=80;腹部 n=52)作为对照组。

结果

观察者间图像质量评分的一致性良好(k=0.74)。高螺距 CT 的主观图像质量在所有患者中均具有诊断价值(胸部中位数评分:2,腹部中位数评分:2)。高螺距 CT 的图像噪声与常规螺距胸部 CT 相当(p=0.32),但在腹部增加(p<0.0001)。高螺距 CT 的辐射剂量为 4.4±0.9 mSv(胸部)和 6.5±1.2 mSv(腹部)。与常规螺距 CT 相比,这些值显著降低(胸部:5.5±1.2 mSv;腹部:11.3±3.8 mSv)。

结论

基于技术背景,高螺距双源 CT 可作为低辐射剂量常规胸部和腹部 CT 的替代扫描模式。