Taheriazam Afshin, Mohajer Mohammad Azizbaig, Aboulghasemian Mansoour, Hajipour Babak
Baqiyatallah Hospital, Baqiyatallah University of Medical Sciences, Iran.
Orthopedics. 2012 Jan 16;35(1):e91-3. doi: 10.3928/01477447-20111122-33.
The fracture rate of third-generation ceramic liners is greatly reduced compared with first- and second-generation liners because of improvements in the design and manufacturing process. Fractures of the alumina-bearing couple are rare for the same reason.This article describes a case of a fracture of an alumina-bearing couple delta ceramic liner without trauma history that was treated with ceramic-on-polyethylene revision total hip arthroplasty. A 57-year-old man was admitted to the hip ward because of an alumina-bearing couple delta ceramic liner fracture. He underwent hip replacement by anterior approach 18 months previously in the same center because of left hip primary osteoarthritis. He received a 54×36-mm modular press-fit cup ceramic alumina-bearing couple delta insert. Probable causes of such fractures are manufacture production failure and edge loading based on cup inclination, but in our patient, inacceptable range of motion, failure of the locking mechanism during implantation insertion, or cracking were possible causes of fracture.Although the fracture rate of third-generation alumina-bearing couples is low, we believe that it may not be possible to eliminate the actual risk of alumina head fracture. Patients should be informed about the potential for this complication before receiving an alumina-bearing couple.
与第一代和第二代陶瓷内衬相比,由于设计和制造工艺的改进,第三代陶瓷内衬的骨折率大幅降低。出于同样的原因,含氧化铝配对部件的骨折很少见。本文描述了一例无外伤史的含氧化铝配对部件delta陶瓷内衬骨折病例,该病例采用陶瓷对聚乙烯翻修全髋关节置换术进行治疗。一名57岁男性因含氧化铝配对部件delta陶瓷内衬骨折入住髋关节病房。他18个月前因左髋原发性骨关节炎在同一中心接受了前路髋关节置换术。他接受了一个54×36毫米的模块化压配髋臼杯陶瓷含氧化铝配对部件delta内衬。这种骨折的可能原因是制造生产故障以及基于髋臼倾斜度的边缘负荷,但在我们的患者中,运动范围不可接受、植入插入过程中锁定机制故障或出现裂纹也可能是骨折的原因。尽管第三代含氧化铝配对部件的骨折率较低,但我们认为可能无法消除氧化铝股骨头骨折的实际风险。在患者接受含氧化铝配对部件之前,应告知其这种并发症的可能性。