Department of Obstetrics and Gynecology, West China Second University Hospital, Sichuan University, Chengdu 610041, China.
Arch Gynecol Obstet. 2012 Jun;285(6):1637-41. doi: 10.1007/s00404-011-2207-0. Epub 2012 Jan 11.
It is to compare the thermal damage on myometrium tissue caused by five electrosurgical instruments, including monopolar forceps, bipolar forceps, PK scalpel, Ligasure and BiClamp.
Normal myometrium in vitro was collected and electric coagulation was conducted with five electrosurgical instruments under corresponding powers. The zones of thermal injury (ZTI) in each coagulation sites were examined histologically, while the width and depth of thermal damage were measured.
As for myometrium, the thermal damage is rather small in the horizontal and vertical directions when using BiClamp and PK scalpel. Ligasure places larger range of thermal damage in horizontal direction with little depth in vertical direction, which is rather safe when acting on uterine surface. Electrocoagulation was conducted with monopolar (the power is 55 W) and bipolar forceps (the power is 40 W) continuously for 3 s, whose thermal damage range is fairly safe to corpus uteri wall and fundus uteri.
比较 5 种电外科器械(单极抓钳、双极抓钳、PK 刀、Ligasure 和 BiClamp)对子宫肌组织热损伤的影响。
取正常离体子宫肌组织,在相应功率下使用 5 种电外科器械进行电凝。组织学检查各电凝部位的热损伤区(ZTI),测量热损伤的宽度和深度。
对于子宫肌组织,使用 BiClamp 和 PK 刀时,水平和垂直方向的热损伤范围较小。Ligasure 在水平方向上造成较大范围的热损伤,垂直方向上的深度较小,在子宫表面操作时较为安全。使用单极(功率 55 W)和双极抓钳(功率 40 W)连续电凝 3 s,其热损伤范围对子宫壁和宫底较为安全。