Chan Maria F, Song Yulin, Dauer Lawrence T, Li Jingdong, Huang David, Burman Chandra
Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, NY, USA.
Med Dosim. 2012 Autumn;37(3):334-8. doi: 10.1016/j.meddos.2011.11.007. Epub 2012 Jan 31.
The purpose of this work was to determine the relative sensitivity of skin QED diodes, optically stimulated luminescent dosimeters (OSLDs) (microStar™ DOT, Landauer), and LiF thermoluminescent dosimeters (TLDs) as a function of distance from a photon beam field edge when applied to measure dose at out-of-field points. These detectors have been used to estimate radiation dose to patients' implantable cardioverter-defibrillators (ICDs) located outside the treatment field. The ICDs have a thin outer case made of 0.4- to 0.6-mm-thick titanium (∼2.4-mm tissue equivalent). A 5-mm bolus, being the equivalent depth of the devices under the patient's skin, was placed over the ICDs. Response per unit absorbed dose-to-water was measured for each of the dosimeters with and without bolus on the beam central axis (CAX) and at a distance up to 20 cm from the CAX. Doses were measured with an ionization chamber at various depths for 6- and 15-MV x-rays on a Varian Clinac-iX linear accelerator. Relative sensitivity of the detectors was determined as the ratio of the sensitivity at each off-axis distance to that at the CAX. The detector sensitivity as a function of the distance from the field edge changed by ± 3% (1-11%) for LiF TLD-700, decreased by 10% (5-21%) for OSLD, and increased by 16% (11-19%) for the skin QED diode (Sun Nuclear Corp.) at the equivalent depth of 5 mm for 6- or 15-MV photon energies. Our results showed that the use of bolus with proper thickness (i.e., ∼d(max) of the photon energy) on the top of the ICD would reduce the scattered dose to a lower level. Dosimeters should be calibrated out-of-field and preferably with bolus equal in thickness to the depth of interest. This can be readily performed in clinic.
这项工作的目的是确定皮肤量子增强探测器(QED)二极管、光激励发光剂量计(OSLD)(MicroStar™ DOT,Landauer公司)和氟化锂热释光剂量计(TLD)在应用于测量射野外点剂量时,作为距光子束野边缘距离的函数的相对灵敏度。这些探测器已被用于估计位于治疗野外的患者植入式心脏复律除颤器(ICD)所接受的辐射剂量。ICD有一个由0.4至0.6毫米厚的钛制成的薄外壳(约2.4毫米组织等效物)。在ICD上方放置一个5毫米厚的等效物,其为设备在患者皮肤下的等效深度。在束流中心轴(CAX)上以及距CAX最远20厘米处,分别测量了有和没有等效物时每个剂量计每单位吸收剂量对水的响应。在瓦里安Clinac-iX直线加速器上,使用电离室在不同深度测量了6兆伏和15兆伏X射线的剂量。探测器的相对灵敏度被确定为每个离轴距离处的灵敏度与CAX处灵敏度的比值。对于6兆伏或15兆伏光子能量,在5毫米等效深度处,LiF TLD-700探测器灵敏度随距野边缘距离的变化为±3%(1 - 11%),OSLD降低了10%(5 - 21%),皮肤QED二极管(太阳核公司)增加了16%(11 - 19%)。我们的结果表明,在ICD顶部使用厚度合适(即光子能量的约d(max))的等效物可将散射剂量降低到较低水平。剂量计应在射野外进行校准,最好使用厚度与感兴趣深度相等的等效物。这在临床中很容易做到。