Int J Oral Maxillofac Implants. 2012 Jan-Feb;27(1):128-37.
To evaluate the accuracy of a newly developed open-source system for three-dimensional dental implant planning and fully guided dental implant placement.
Forty-eight implant positions were planned for eight partially edentulous anatomical patient equivalent models with an open-source implant planning system on the basis of cone beam computed tomography (CBCT) scans. The virtual software planning leads to an output of four different coordinates for each implant position; with these, surgical guides were manufactured using a coordinate-transfer apparatus. During the surgical simulation, drills and implants were fully guided as they were inserted by means of the harmonized components of the vendor's sleeve-in-sleeve system. After follow-up CBCT investigation and reference marker-based software registration, linear horizontal, vertical, and maximal 3D deviations, as well as angular deviations, between the virtual planning data and the surgical results were calculated.
The mean three-dimensional deviation values for the final implant positions were 671 μm (95% confidence interval [CI] 452 to 891 μm) at the implant base and 808 μm (95% CI 646 to 971 μm) at the implant tip. Mean vertical deviations were 273 μm (95% CI 200 to 345 μm). Mean angular deviations of 1.9 degrees (95% CI 1.4 to 2.4 degrees) were measured.
The open-source implant planning system described in this study demonstrated a level of accuracy that is equal or superior to most descriptions of the literature on computer-aided implant dentistry, allows for predictable implant positioning, and has the potential to reduce postoperative impairment versus conventional implant insertion.
评估一种新开发的开源系统在三维牙科种植体规划和全导向牙科种植体植入中的准确性。
根据锥形束计算机断层扫描(CBCT)扫描,使用开源种植体规划系统对 8 个部分无牙解剖学患者等效模型的 48 个种植体位置进行规划。虚拟软件规划为每个种植体位置生成四个不同的坐标;使用坐标转换设备制造这些坐标的手术导板。在手术模拟过程中,通过供应商的套管-套管系统的协调组件,钻头和种植体被完全引导插入。在后续的 CBCT 检查和基于参考标记的软件注册后,计算了虚拟规划数据与手术结果之间的线性水平、垂直和最大三维偏差以及角度偏差。
最终种植体位置的三维偏差平均值为种植体基底处 671μm(95%置信区间[CI]为 452 至 891μm)和种植体尖端处 808μm(95%CI 为 646 至 971μm)。平均垂直偏差为 273μm(95%CI 为 200 至 345μm)。测量的平均角度偏差为 1.9 度(95%CI 为 1.4 至 2.4 度)。
本研究中描述的开源种植体规划系统表现出与计算机辅助种植牙文献中大多数描述相当或更高的准确性,允许可预测的种植体定位,并有可能减少术后损伤,而不是传统的种植体插入。