Comparison between a high-resolution single-stage Orbitrap and a triple quadrupole mass spectrometer for quantitative analyses of drugs.

机构信息

Clinical Chemistry, University Hospital of Lausanne; CHUV, 1011, Lausanne, Switzerland.

出版信息

Rapid Commun Mass Spectrom. 2012 Mar 15;26(5):499-509. doi: 10.1002/rcm.6121.

Abstract

The capabilities of a high-resolution (HR), accurate mass spectrometer (Exactive-MS) operating in full scan MS mode was investigated for the quantitative LC/MS analysis of drugs in patients' plasma samples. A mass resolution of 50,000 (FWHM) at m/z 200 and a mass extracted window of 5 ppm around the theoretical m/z of each analyte were used to construct chromatograms for quantitation. The quantitative performance of the Exactive-MS was compared with that of a triple quadrupole mass spectrometer (TQ-MS), TSQ Quantum Discovery or Quantum Ultra, operating in the conventional selected reaction monitoring (SRM) mode. The study consisted of 17 therapeutic drugs including 8 antifungal agents (anidulafungin, caspofungin, fluconazole, itraconazole, hydroxyitraconazole posaconazole, voriconazole and voriconazole-N-oxide), 4 immunosuppressants (ciclosporine, everolimus, sirolimus and tacrolimus) and 5 protein kinase inhibitors (dasatinib, imatinib, nilotinib, sorafenib and sunitinib). The quantitative results obtained with HR-MS acquisition show comparable detection specificity, assay precision, accuracy, linearity and sensitivity to SRM acquisition. Importantly, HR-MS offers several benefits over TQ-MS technology: absence of SRM optimization, time saving when changing the analysis from one MS to another, more complete information of what is in the samples and easier troubleshooting. Our work demonstrates that U/HPLC coupled to Exactive HR-MS delivers comparable results to TQ-MS in routine quantitative drug analyses. Considering the advantages of HR-MS, these results suggest that, in the near future, there should be a shift in how routine quantitative analyses of small molecules, particularly for therapeutic drugs, are performed.

摘要

高分辨率(HR)、精确质量质谱仪(Exactive-MS)在全扫描 MS 模式下的性能,用于患者血浆样品中药物的 LC/MS 定量分析。在 m/z 200 处的质量分辨率为 50,000(半峰全宽),在每个分析物的理论 m/z 周围的质量提取窗口为 5 ppm,用于构建定量色谱图。Exactive-MS 的定量性能与三重四极杆质谱仪(TQ-MS)、TSQ Quantum Discovery 或 Quantum Ultra 进行比较,后者以传统的选择反应监测(SRM)模式运行。该研究包括 17 种治疗药物,包括 8 种抗真菌药物(阿尼达卢凡、卡泊芬净、氟康唑、伊曲康唑、羟基伊曲康唑泊沙康唑、伏立康唑和伏立康唑-N-氧化物)、4 种免疫抑制剂(环孢菌素、依维莫司、西罗莫司和他克莫司)和 5 种蛋白激酶抑制剂(达沙替尼、伊马替尼、尼洛替尼、索拉非尼和舒尼替尼)。使用 HR-MS 采集获得的定量结果显示,与 SRM 采集相比,具有相当的检测特异性、分析精密度、准确度、线性和灵敏度。重要的是,HR-MS 相对于 TQ-MS 技术具有以下优势:无需进行 SRM 优化,在从一种 MS 切换到另一种 MS 进行分析时节省时间,提供样品中更多的信息,以及更易于进行故障排除。我们的工作表明,U/HPLC 与 Exactive HR-MS 相结合,在常规定量药物分析中可获得与 TQ-MS 相当的结果。考虑到 HR-MS 的优势,这些结果表明,在不久的将来,小分子的常规定量分析,特别是治疗药物的定量分析,应该采用新的方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验