Reproducibility of an optical measurement system for the clinical evaluation of active knee rotation in weight-bearing, healthy subjects.

机构信息

Lyon University, 69000, Lyon, France.

出版信息

Orthop Traumatol Surg Res. 2012 Apr;98(2):159-66. doi: 10.1016/j.otsr.2011.08.017. Epub 2012 Feb 14.

Abstract

INTRODUCTION

A knee is typically evaluated passively by a clinician during an office visit, without using dedicated measurement tools. When the knee is evaluated with the patient standing and actively participating in the movement, the results will differ than when the knee is passively moved through its range-of-motion by the surgeon. If a precise measurement system was available, it could provide additional information to the clinician during this evaluation.

HYPOTHESIS

The goal of this study was to verify the reproducibility of a fast, flexible optical measurement system to measure rotational knee laxity during weight-bearing.

MATERIAL AND METHODS

Two passive reflective targets were placed on the legs of 11 subjects to monitor femur and tibia displacements in three dimensions. Subjects performed internal and external rotation movements with the knee extended or flexed 30°. During each movement, seven variables were measured: internal rotation, external rotation and overall laxity in extension and 30° flexion, along with neutral rotation value in 30° flexion. Measurement accuracy was also assessed and the right and left knees were compared. Reproducibility was assessed over two measurements sessions.

RESULTS

The calculated intra-class correlation coefficient (ICC) for reproducibility was above 0.9 for five of the seven variables measured. The calculated ICC for the right/left comparison was above 0.75 for five of the seven variables measured.

DISCUSSION

These results confirmed that the proposed system provides reproducible measurements. Our right/left comparison results were consistent with the published literature. This system is fast, reproducible and flexible, which makes it suitable for assessing various weight-bearing movements during clinical evaluations.

LEVEL OF EVIDENCE

Level III, experimental study.

摘要

简介

在就诊期间,临床医生通常通过被动的方式评估膝关节,而不使用专用的测量工具。当患者站立并主动参与运动时,对膝关节进行评估,与通过外科医生被动地移动膝关节使其在活动范围内移动时的评估结果会有所不同。如果有一个精确的测量系统可用,它可以在评估过程中为临床医生提供额外的信息。

假设

本研究的目的是验证一种快速、灵活的光学测量系统在负重时测量膝关节旋转松弛度的可重复性。

材料与方法

在 11 名受试者的腿部放置两个被动反射性标记物,以监测股骨和胫骨在三个维度上的位移。受试者在膝关节伸直或弯曲 30°的情况下进行内旋和外旋运动。在每次运动中,测量七个变量:伸直和 30°弯曲时的内旋、外旋和整体松弛度,以及 30°弯曲时中立旋转值。还评估了测量精度,并比较了右膝和左膝。在两次测量中评估了可重复性。

结果

对于测量的七个变量中的五个,计算出的组内相关系数(ICC)的可重复性大于 0.9。对于测量的七个变量中的五个,计算出的右/左比较的 ICC 大于 0.75。

讨论

这些结果证实了所提出的系统提供了可重复的测量结果。我们的右/左比较结果与已发表的文献一致。该系统快速、可重复且灵活,非常适合在临床评估中评估各种负重运动。

证据水平

三级,实验研究。