评估带有集成动脉过滤器的Quadrox-i和Capiox FX新生儿氧合器在模拟体外循环期间消除气态微栓子并保持血流动力学特性的能力。

Evaluation of Quadrox-i and Capiox FX neonatal oxygenators with integrated arterial filters in eliminating gaseous microemboli and retaining hemodynamic properties during simulated cardiopulmonary bypass.

作者信息

Lin J, Dogal N M, Mathis R K, Qiu F, Kunselman A, Ündar A

机构信息

Department of Pediatrics, Penn State Milton S. Hershey Medical Center, Penn State Hershey College of Medicine, Penn State Hershey Children's Hospital, Hershey, Pennsylvania 17033-0850, USA.

出版信息

Perfusion. 2012 May;27(3):235-43. doi: 10.1177/0267659112438932. Epub 2012 Feb 15.

Abstract

Perfusion quality during cardiopulmonary bypass (CPB) procedures can contribute to postoperative neurological complications and influence patient recovery and outcome. Gaseous microemboli generated in the circuit and hemodynamic properties of blood reaching the patient can be monitored during CPB to optimize perfusion. Oxygenators that oxygenate the blood during CPB can significantly influence the quality of blood reaching the patient by their manufacturing designs. New hollow-fiber membrane oxygenators are developed with integrated arterial filters to reduce priming volume and eliminate a separate arterial filter in the circuit. To evaluate the performance of these new oxygenators, we used a simulated model to compare the Quadrox-i Neonatal and the Capiox Baby FX05 neonatal oxygenators and to provide a review of these oxygenators with their respective counterparts which have separate arterial filters. We found that microemboli counts for the new Quadrox-i and Capiox FX05 oxygenators are similar in the arterial line, but different across the oxygenator for all experimental conditions. The arterial purge line diverting blood from the patient reduces microemboli count for the Capiox FX05, but is inconsistent for the Quadrox-i Neonatal. While hemodynamic energy delivered to the patient is similar for both oxygenators, shunted blood flow for the Quadrox-i Neonatal oxygenator is three times higher than the Capiox FX05 (103.6 mL/min vs 33.0 mL/min at 400 mL/min and 35°C) (p<0.001).

摘要

体外循环(CPB)手术期间的灌注质量可能导致术后神经并发症,并影响患者的恢复和预后。在CPB期间可以监测体外循环中产生的气态微栓子以及到达患者体内的血液的血流动力学特性,以优化灌注。在CPB期间对血液进行氧合的氧合器,其制造设计会显著影响到达患者体内的血液质量。新型中空纤维膜氧合器集成了动脉滤器,以减少预充量并消除体外循环中的单独动脉滤器。为了评估这些新型氧合器的性能,我们使用模拟模型比较了Quadrox-i新生儿氧合器和Capiox Baby FX05新生儿氧合器,并对这些氧合器及其具有单独动脉滤器的同类产品进行了综述。我们发现,新型Quadrox-i和Capiox FX05氧合器在动脉管路中的微栓子计数相似,但在所有实验条件下,氧合器内的微栓子计数有所不同。将血液从患者处引出的动脉冲洗管路可减少Capiox FX05的微栓子计数,但对Quadrox-i新生儿氧合器而言并不一致。虽然两种氧合器输送给患者的血流动力学能量相似,但Quadrox-i新生儿氧合器的分流血流量比Capiox FX05高3倍(在400 mL/min和35°C时分别为103.6 mL/min和33.0 mL/min)(p<0.001)。