Suppr超能文献

开发一种使用离子对紊流在线萃取的无血浆间甲肾上腺素的液相色谱-串联质谱检测法。

Development of a liquid chromatography-tandem mass spectrometry method for plasma-free metanephrines with ion-pairing turbulent flow online extraction.

机构信息

ThermoFisher Scientific, San Jose, CA 95134, USA.

出版信息

Anal Bioanal Chem. 2012 Mar;402(9):3003-10. doi: 10.1007/s00216-012-5768-2. Epub 2012 Feb 17.

Abstract

Liquid chromatography-tandem mass spectrometry has become the preferred technology to measure unconjugated metanephrine and normetanephrine in plasma because of its high sensitivity and specificity over immunoassay and gas chromatography-mass spectrometry. In our earlier study, plasma metanephrines were extracted with offline ion-pairing solid-phase extraction and quantified by liquid chromatography-tandem mass spectrometry with porous graphitic carbon column based chromatography. In this study, we aim to automate the sample preparation with turbulent flow online extraction technology and maintain or improve the analytical performance previously achieved from the offline approach. The online extraction was done with a mixed-mode cation exchange turbulent flow chromatography column assisted with ion-pairing reagent and porous graphitic column was used for chromatographic separation. The total online extraction and analytical LC runtime was 12 min. This method was linear from 6.3 to 455.4 pg/mL for metanephrine; 12.6 to 954.5 pg/mL for normetanephrine with an accuracy of 80.6% to 93.5% and 80.9% to 101.7%, respectively. The lower limit of quantitation was 6.3 pg/mL for metanephrine and 12.6 pg/mL for normetanephrine. Inter-assay and intra-assay precision for metanephrine and normetanephrine at low and high concentration levels ranged from 2.0% to 10.5%. In conclusion, we have developed a fast and sensitive automated online turbulent flow extraction method for the quantitative analysis of plasma metanephrines. Ion-pairing reagent was necessary for the success of this method.

摘要

液相色谱-串联质谱法因其灵敏度和特异性高于免疫测定法和气相色谱-质谱法,已成为测量血浆中非结合性甲肾上腺素和去甲肾上腺素的首选技术。在我们之前的研究中,使用离线离子对固相萃取提取血浆甲肾上腺素,并通过基于多孔石墨碳柱的色谱的液相色谱-串联质谱法定量。在这项研究中,我们旨在使用湍流在线提取技术实现样品制备自动化,并保持或提高以前从离线方法获得的分析性能。在线提取是使用混合模式阳离子交换湍流色谱柱辅助离子对试剂和多孔石墨柱进行的色谱分离。总在线提取和分析 LC 运行时间为 12 分钟。该方法对于甲肾上腺素呈线性,范围为 6.3 至 455.4 pg/mL;对于去甲肾上腺素呈线性,范围为 12.6 至 954.5 pg/mL,准确度为 80.6%至 93.5%,80.9%至 101.7%。甲肾上腺素的定量下限为 6.3 pg/mL,去甲肾上腺素的定量下限为 12.6 pg/mL。甲肾上腺素和去甲肾上腺素在低浓度和高浓度水平下的批内和批间精密度范围为 2.0%至 10.5%。总之,我们已经开发出一种快速灵敏的自动化在线湍流提取方法,用于定量分析血浆中甲肾上腺素。离子对试剂对于该方法的成功是必要的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验