Dance D R
Department of Physics, Royal Marsden Hospital, London, UK.
Phys Med Biol. 1990 Sep;35(9):1211-9. doi: 10.1088/0031-9155/35/9/002.
The IPSM report on the commissioning and routine testing of mammographic x-ray systems recommends that breast dose be specified as the mean dose to the glandular tissues within the breast and gives the size and compositions of a standard breast phantom for the comparison of doses. The dose to this standard breast phantom can be determined by measuring the incident air kerma to a Perspex phantom and applying appropriate multiplicative conversion factors. These conversion factors have been evaluated by Monte Carlo calculations for a wide range of mammographic x-ray spectra. Some factors are provided for a range of breast thicknesses to supplement existing tabulations. Results are also given for equivalent thicknesses of Perspex and breast tissue.
乳腺X射线系统调试与常规检测的IPSM报告建议,将乳房剂量规定为乳房内腺体组织的平均剂量,并给出用于剂量比较的标准乳房体模的尺寸和组成。通过测量有机玻璃体模的入射空气比释动能并应用适当的乘法转换因子,可以确定该标准乳房体模的剂量。这些转换因子已通过蒙特卡罗计算针对广泛的乳腺X射线光谱进行了评估。提供了一系列乳房厚度的一些因子,以补充现有表格。还给出了有机玻璃和乳腺组织等效厚度的结果。