Eur J Radiol. 2012 Nov;81(11):3250-3. doi: 10.1016/j.ejrad.2012.02.001. Epub 2012 Feb 26.
The purpose of this study was to evaluate the correlation between bolus-tracking parameters and heart rate (HR) with attenuation of the ascending aorta and create a linear regression model for predicting coronary attenuation in coronary computed tomography angiography (CCTA).
A total of 50 patients (31 men, 19 women; mean age, 67.2±10.8 y) underwent CCTA using a 320-detector CT scanner. A bolus-tracking scan was performed to optimize the scan timing. The average HR under normal breathing for 10s was recorded just before the bolus-tracking scan started. Attenuation values of the pulmonary artery at 7s (PA7) and 10s (PA10) after the beginning of the injection were recorded during the bolus-tracking scan and the ascending aortic attenuation (CEAAo) was measured during the diagnostic scan.
A positive correlation was observed between PA7 and CEAAo (r=0.41, P=0.003) and PA10 and CEAAo (r=0.66, P<0.0001), and weak negative correlation was observed between HR and CEAAo (r=-0.46, P=0.15). A multivariable linear regression model for predicting CEAAo was evaluated, and the residual error between the predicted and the measured CEAAo was within approximately ±100 HU.
Coronary attenuation could be predicted using HR and pulmonary artery attenuation during the bolus-tracking method.
本研究旨在评估升主动脉衰减与团注追踪参数和心率(HR)之间的相关性,并建立用于预测冠状动脉 CT 血管造影(CCTA)中冠状动脉衰减的线性回归模型。
共 50 例患者(男 31 例,女 19 例;平均年龄 67.2±10.8 岁)接受了使用 320 探测器 CT 扫描仪进行的 CCTA。进行了团注追踪扫描以优化扫描时间。在开始团注追踪扫描之前,记录正常呼吸 10 秒时的平均 HR。在团注追踪扫描期间记录注射开始后 7s(PA7)和 10s(PA10)时肺动脉的衰减值,并在诊断扫描期间测量升主动脉衰减(CEAAo)。
PA7 与 CEAAo 之间存在正相关(r=0.41,P=0.003),PA10 与 CEAAo 之间存在强正相关(r=0.66,P<0.0001),HR 与 CEAAo 之间存在弱负相关(r=-0.46,P=0.15)。评估了用于预测 CEAAo 的多变量线性回归模型,预测的和实测的 CEAAo 之间的残差约为±100 HU。
可以使用团注追踪方法中的 HR 和肺动脉衰减来预测冠状动脉衰减。