Suppr超能文献

[The comparison between dose rates at the interventional reference point of the angiography systems in many facilities].

作者信息

Ishibashi Tooru, Imada Naoyuki, Yamashita Yukari, Asou Hiroya, Matsumoto Yoriaki, Inada Satoshi, Okino Mizuho, Nonaka Haruki, Mizutani Hiroshi

机构信息

Department of Medical Technology, Tsuchiya General Hospital.

出版信息

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2012;68(3):260-8. doi: 10.6009/jjrt.2012_jsrt_68.3.260.

Abstract

The management of the radiation dose is very important in interventional radiology (IVR), especially in percutaneous coronary intervention (PCI). Therefore, we measured entrance surface doses at the interventional reference point of 27 cardiac intervention procedures in 22 cardiac catheterization laboratories around Hiroshima, and compared these doses. Recently, for cardiac interventional radiology, the X-ray machines using flat-panel detectors (FPD) instead of image intensifiers (I.I.) is increasing; 13 systems used FPD and 14 systems used I.I. For fluoroscopy rate, the difference between laboratories was 9 times. For cineangiography rate, the difference between laboratories was 7 times. In addition, between both devices, the I.I. group is bigger than the FPD group. When comparing by the same condition, for the dose at the interventional reference point, no significant difference was detected between the FPD group and the I.I. group. This study shows that FPD is not available for reducing the radiation dose simply. Therefore, it is necessary that we think of the balance with image quality and radiation dose. The optimization of the devices and cardiac intervention procedures becomes very important.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验