Zhu Xi, Zhen Yu, Mi Tie-Zhu, Chi Zhen-Ming, Xu Xiao-Chun
College of Marine Life Sciences, Ocean University of China, Qingdao 266003, China.
Huan Jing Ke Xue. 2012 Jan;33(1):222-7.
To develop a new technique, Electrochemiluminescence-Molecular Probe (ECL-MP), for harmful algae detection, a highly sensitive electrochemiluminescence (ECL) detection system was set up based on the principle of ECL and related literature. The optimization tests were carried out, including the impact of voltage and current, the concentration of tripropylamine (TPrA) and the pH of phosphate buffer (PBS). The determination limit of Ru (bpy)3Cl2 x 6H2O was 10(-11) mol x L(-1) and the detection range was from 10(-9) - 10(-5) mol x L(-1) as well as the detection amount of substance was in the range from 0.4 pmol-4 nmol in the optimal reaction conditions with voltage of 1.0 V, current of 1.0 mA, TPrA concentration of 1.5 mol x L(-1) and pH of 7.4. It was proved by facts that this ECL analyzer was stable and highly sensitive, which was helpful in establishment of ECL-MP.
为开发一种用于有害藻类检测的新技术——电化学发光分子探针(ECL-MP),基于电化学发光原理并参考相关文献建立了一种高灵敏度的电化学发光(ECL)检测系统。进行了优化试验,包括电压和电流的影响、三丙胺(TPrA)浓度以及磷酸盐缓冲液(PBS)的pH值。在最佳反应条件下,即电压为1.0 V、电流为1.0 mA、TPrA浓度为1.5 mol·L⁻¹且pH值为7.4时,Ru(bpy)₃Cl₂·6H₂O的测定限为10⁻¹¹ mol·L⁻¹,检测范围为10⁻⁹ - 10⁻⁵ mol·L⁻¹,物质检测量范围为0.4 pmol - 4 nmol。事实证明,该ECL分析仪稳定且灵敏度高,有助于建立ECL-MP。