Persson A M, Sokolowski A, Pettersson C
Division of Analytical Pharmaceutical Chemistry, Uppsala University, BMC, SE-751 23 Uppsala, Sweden.
Drug Discov Ther. 2009 Jun;3(3):104-13.
A correlation of the logarithmic values of the in vitro dissolution rate, G, and the apparent solubility, S, was evaluated in phosphate and ammonium acetate buffer at an initial pH of 7. The dissolution rates were determined with a newly designed and build miniaturized rotating disk equipment, as well as with a traditional rotating disk apparatus. The two apparatuses gave the same correlation pattern of logG and logS. Thirteen diverse drug substances from all of the classes in the Biopharmaceutics Classification System (BCS) were used for the correlation in the phosphate buffer system, with the results from the miniaturized apparatus only. A coefficient of determination, R2, of 0.982 was found if bases formulated as hydrochloride salts were excluded in the correlation. The miniaturized equipment is used for rapid screening of the dissolution rate, approximately 10 min for one run, and consumes small amounts of substance (about 5 mg) and dissolution media. All quantifications were performed by using reversed phase high-performance liquid chromatography (RPHPLC) with a diode array detector (DAD), integrated with the miniaturized rotating disk equipment.
在初始pH值为7的磷酸盐和醋酸铵缓冲液中,评估了体外溶出速率G的对数值与表观溶解度S的对数值之间的相关性。溶出速率通过新设计制造的小型旋转圆盘设备以及传统旋转圆盘装置进行测定。这两种装置给出了相同的logG和logS相关性模式。在磷酸盐缓冲系统中,使用了生物药剂学分类系统(BCS)中所有类别的13种不同药物进行相关性研究,仅采用小型设备的结果。如果在相关性研究中排除以盐酸盐形式配制的碱,则决定系数R2为0.982。小型设备用于快速筛选溶出速率,一次运行约10分钟,消耗少量物质(约5毫克)和溶出介质。所有定量分析均采用配备二极管阵列检测器(DAD)的反相高效液相色谱(RPHPLC)进行,该检测器与小型旋转圆盘设备集成在一起。