二维经阴道超声测量子宫下段厚度视频序列的可靠性。

Reliability of two-dimensional transvaginal sonographic measurement of lower uterine segment thickness using video sequences.

机构信息

Department of Social and Preventive Medicine, Faculty of Medicine, Université Laval, Québec, Canada.

出版信息

Am J Perinatol. 2012 Aug;29(7):527-32. doi: 10.1055/s-0032-1310524. Epub 2012 Apr 11.

Abstract

OBJECTIVES

To report the intra- and interobserver reliability of measurement of the lower uterine segment (LUS) thickness using transvaginal sonographic videos.

METHODS

A prospective study of 60 women with previous, low-transverse cesarean undergoing LUS examination (36 to 39 weeks) was performed. Two observers independently measured full LUS thickness using transvaginal sonography. A video of the LUS was recorded and analyzed more than 2 months later by both observers. Intra- and interobserver reliability was assessed with median absolute differences and interquartile range (IQR), nonparametric limits of agreement, intraclass correlation coefficients (ICC) with 95% confidence interval (95% CI), and kappa coefficients.

RESULTS

Median full LUS thickness was 3.6 mm (range: 0.9 to 8.0 mm). Intraobserver repeatability was excellent (median difference: 0.2 mm, IQR: 0.1 to 0.4; ICC: 0.94, 95% CI: 0.90 to 0.96; kappa: 1.00). Interobserver (median difference: 0.3 mm, IQR: 0.2 to 1.3; ICC: 0.91, 95% CI: 0.86 to 0.95; kappa: 0.76, 95% CI: 0.54 to 0.98) and intermethod reproducibility (median difference: 0.4 mm, IQR: 0.2 to 0.8; ICC: 0.82, 95% CI: 0.72 to 0.89; kappa: 0.69, 95% CI: 0.43 to 0.94) were good. However, both interobserver and intermethod reproducibility were improved when LUS thickness was below 3 mm.

CONCLUSION

Full LUS thickness measured from transvaginal sonographic videos has excellent intra- and interobserver reproducibility and good reproducibility with live transvaginal ultrasound.

摘要

目的

报告使用经阴道超声视频测量子宫下段(LUS)厚度的观察者内和观察者间可靠性。

方法

对 60 名经阴道超声检查的既往低横位剖宫产妇女(36 至 39 周)进行了前瞻性研究。两名观察者独立使用经阴道超声测量完整的 LUS 厚度。记录 LUS 的视频,并由两名观察者在 2 个月后以上进行分析。使用中位数绝对差异和四分位间距(IQR)、非参数协议界限、95%置信区间(95%CI)内的组内相关系数(ICC)和kappa 系数评估观察者内和观察者间的可靠性。

结果

中位完整 LUS 厚度为 3.6 毫米(范围:0.9 至 8.0 毫米)。观察者内重复性极好(中位数差异:0.2 毫米,IQR:0.1 至 0.4;ICC:0.94,95%CI:0.90 至 0.96;kappa:1.00)。观察者间(中位数差异:0.3 毫米,IQR:0.2 至 1.3;ICC:0.91,95%CI:0.86 至 0.95;kappa:0.76,95%CI:0.54 至 0.98)和间方法重复性(中位数差异:0.4 毫米,IQR:0.2 至 0.8;ICC:0.82,95%CI:0.72 至 0.89;kappa:0.69,95%CI:0.43 至 0.94)良好。然而,当 LUS 厚度低于 3 毫米时,观察者间和间方法的重复性均得到改善。

结论

从经阴道超声视频测量的完整 LUS 厚度具有极好的观察者内和观察者间可重复性,并且与实时经阴道超声具有良好的可重复性。