Oropesa P, Moreno Y, Serra R A, Hernández A T
Centro de Isótopos (CENTIS). Avenue Monumental y Carretera La Rada Km 3 1/2. Guanabacoa, La Habana 11100, Cuba.
Appl Radiat Isot. 2012 Sep;70(9):2251-4. doi: 10.1016/j.apradiso.2012.02.108. Epub 2012 Mar 5.
The national traceability chain for (131)I activity measurements performed in nuclear medicine in Cuba is described. At the highest (primary) level, liquid scintillation counting employing the CIEMAT/NIST method is used; at the secondary level, a secondary standard radionuclide calibrator is utilized that allows for a quick and simple transference of the measurement unit to the tertiary level of end-users' instruments. The equivalence of Cuban standards and the assessment of measurement uncertainties at the end-user level are determined through the results of measurement comparisons.