Faber J, Gam A, Siersbaek-Nielsen K
Department of Endocrinology E, Frederiksberg Hospital, Denmark.
Acta Endocrinol (Copenh). 1990 Nov;123(5):535-40. doi: 10.1530/acta.0.1230535.
Most serum TSH assays have a working sensitivity (i.e. the lowest TSH value with an inter-assay coefficient of variation below 10%) around 0.15-0.4 mU/l, which also is the critical area for cut off for further thyroid profile testing when serum TSH is used as a "first line test". A new assay (BeriLux hTSH) based on chemiluminescence was evaluated, and demonstrated a theoretical sensitivity (mean + 2 SD of the zero standard) of 0.005 mU/l and a working sensitivity as low as 0.04 mU/l. Reference range was 0.18-2.60 mU/l (N = 33). Sixty-eight percent (13/19) of hyperthyroid patients had serum TSH less than 0.005 mU/l, all had serum TSH less than 0.037 mU/l. We studied two groups of patients with normal free T4 and T3 indices but serum TSH less than 0.15 mU/l as measured by an immunoradiometric assay. Thirty-five percent (7/20) of patients with nontoxic goitre and 5% (1/20) of L-T4 treated patients had serum TSH less than 0.005 mU/l; and in 60% (11/20) and 30% (6/20), respectively, the levels overlapped the total range for hyperthyroidism. Serum levels of sex hormone-binding globulin expressed as percent of the reference median for the relevant sex (SHBG%) were elevated in both groups of patients (p less than 0.01). Approximately 50% of those with serum TSH overlapping the hyperthyroid range had serum SHBG% levels above reference range. In conclusion, this assay seems superior to most previously reported assays concerning working sensitivity, but it still leaves us with a group of clinically euthyroid patients who has unmeasurably low serum TSH.(ABSTRACT TRUNCATED AT 250 WORDS)
大多数血清促甲状腺激素(TSH)检测方法的工作灵敏度(即批间变异系数低于10%时的最低TSH值)约为0.15 - 0.4 mU/l,当血清TSH用作“一线检测”时,这也是进一步进行甲状腺功能检测的临界截断区域。对一种基于化学发光的新检测方法(BeriLux hTSH)进行了评估,其理论灵敏度(零标准品的均值 + 2标准差)为0.005 mU/l,工作灵敏度低至0.04 mU/l。参考范围为0.18 - 2.60 mU/l(N = 33)。68%(13/19)的甲状腺功能亢进患者血清TSH低于0.005 mU/l,所有患者血清TSH均低于0.037 mU/l。我们研究了两组游离T4和T3指数正常但免疫放射分析测得血清TSH低于0.15 mU/l的患者。35%(7/20)的非毒性甲状腺肿患者和5%(1/20)接受左甲状腺素(L-T4)治疗的患者血清TSH低于0.005 mU/l;分别有60%(11/20)和30%(6/20)的患者其TSH水平与甲状腺功能亢进的总范围重叠。两组患者中,以相关性别参考中位数百分比表示的性激素结合球蛋白(SHBG%)血清水平均升高(p < 0.01)。血清TSH与甲状腺功能亢进范围重叠的患者中,约50%的人血清SHBG%水平高于参考范围。总之,该检测方法在工作灵敏度方面似乎优于大多数先前报道的检测方法,但仍有一组临床甲状腺功能正常的患者血清TSH低至无法测量。(摘要截短至250字)