The use of inflated bags or cuffs to measure grip strength is now a well established technique. The method does have a number of problems. Bags of different diameter and volume were seen to give statistically significantly different pressure readings when squeezed by the same subjects. Different initial pressures (from 20 mmHg to 60 mmHg) also gave significantly different results both in laboratory tests on a materials testing machine and when patients with rheumatoid arthritis squeezed the bag. The technique of squeezing also affected the results. Despite the intrinsic drawbacks of the system, it is likely to remain in general use because of familiarity and convenience. We recommend the minimum details required when pneumodynamometer-derived data are published.

摘要

使用充气袋或袖带来测量握力现在是一种成熟的技术。该方法确实存在一些问题。当由相同受试者挤压时,不同直径和体积的袋子会给出在统计学上有显著差异的压力读数。不同的初始压力(从20毫米汞柱到60毫米汞柱)在材料试验机上的实验室测试中以及类风湿性关节炎患者挤压袋子时也会产生显著不同的结果。挤压技术也会影响结果。尽管该系统存在内在缺陷,但由于其为人所熟知且使用方便,可能仍会被普遍使用。我们建议在发表气动测力计得出的数据时所需的最少细节。