研究组织体模在超声换能器阵列质量保证中的稳定性测试中的适用性。

Investigation of the suitability of tissue phantoms for testing the constancy of ultrasonic transducer arrays in quality assurance.

机构信息

FB Ingenieur- und Naturwissenschaften, Hochschule Merseburg, Germany.

出版信息

Ultraschall Med. 2012 Jun;33(3):289-94. doi: 10.1055/s-0032-1312808. Epub 2012 May 10.

Abstract

PURPOSE

To ensure high quality ultrasound diagnostics, proper functioning of the devices used is a necessary prerequisite. Ultrasound transducers have proven to be the most failure-prone part of the signal chain. Their technical monitoring is possible in principle with the help of tissue phantoms. The background of the present study is to determine which type of phantoms and which measurement parameters are best suited to a consistency test as part of routine quality assurance of ultrasound imaging systems.

MATERIALS AND METHODS

A classic wire-type phantom (ATS Mod. 539, ATS Labs Bridgeport, USA) and a 3 D cyst phantom (TCC, Timelkam, Austria) were used for the studies and comparative tests were conducted between intact transducers and those in which faults had been simulated. The collected measurement data show a relatively large scatter. Therefore, statistical analysis methods were used, and the discrimination analysis proved to be a useful tool.

RESULTS

Local failures which arise, e. g. due to the breakdown of individual piezoelectric elements or element groups in the transducer array, can be detected with the help of the gray value targets of the ATS phantom, but only in those cases in which the error-affected sound field part actually overlaps with the target under consideration. The TCC phantom is not suitable for the detection of such errors. Global transducer failures, i. e. those that affect the entire array, can even be detected with both types of phantoms.

CONCLUSION

When the emphasis of quality assessment is on the detection of local defects in the array that make up the largest part of the transducer faults, studies with conventional phantoms are only of limited value.

摘要

目的

为了确保高质量的超声诊断,设备的正常运行是必要的前提。超声换能器已被证明是信号链中最容易出现故障的部分。借助组织体模,从原则上讲可以对其进行技术监测。本研究的背景是确定哪种类型的体模和哪种测量参数最适合作为超声成像系统常规质量保证的一致性测试的一部分。

材料和方法

经典的线阵式体模(ATS Mod. 539,ATS Labs Bridgeport,美国)和 3D 囊肿体模(TCC,Timelkam,奥地利)用于研究,并对完好的换能器和模拟故障的换能器进行了对比测试。收集的测量数据显示出较大的分散性。因此,使用了统计分析方法,判别分析被证明是一种有用的工具。

结果

局部故障,例如由于换能器阵元或阵元组中的个别压电元件的击穿而引起的故障,可以借助 ATS 体模的灰度目标进行检测,但仅在受误差影响的声场部分实际上与所考虑的目标重叠的情况下。TCC 体模不适合检测此类误差。全局换能器故障,即影响整个阵列的故障,即使使用这两种类型的体模也可以检测到。

结论

当质量评估的重点是检测构成换能器故障大部分的阵列中的局部缺陷时,使用传统体模的研究仅具有有限的价值。