使用双源并行射频发射技术的 Gd-EOB-DTPA 增强磁共振肝脏成像的图像质量:与用于校正 B1 不均匀性导致的信号丢失的后处理校正方法的比较。

Image quality of Gd-EOB-DTPA-enhanced magnetic resonance imaging of the liver using dual-source parallel radiofrequency transmission technology: Comparison with the post-processing correction method for B1 inhomogeneity-induced signal loss.

机构信息

Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka City, Fukuoka 812-8582, Japan.

出版信息

Eur J Radiol. 2012 Nov;81(11):3035-40. doi: 10.1016/j.ejrad.2012.03.026. Epub 2012 May 9.

Abstract

PURPOSE

To evaluate the efficacy of the dual-source parallel radiofrequency system (DS system) for gadolinium ethoxybenzyl diethylenetriamine pentaacetic acid-enhanced magnetic resonance imaging (Gd-EOB-MRI).

MATERIALS AND METHODS

Twenty-six subjects with liver diseases underwent hepatobiliary phase imaging of Gd-EOB-MRI using either the DS system or a single-source radiofrequency system with or without body-tuned CLEAR, which is a post-processing correction method of B1 inhomogeneity-induced signal loss (SSBTC+ and SSBTC-, respectively). The left and right lobes of the liver were separately assessed. Qualitatively, the homogeneity of signal intensity distribution was scored using a 3-point scale. Quantitatively, lesion-to-liver and spleen-to-liver contrast ratios (CRs) were calculated. The scores and CRs were compared among the three techniques by two-way analysis of variance and Tukey's Honestly Significant Difference post hoc test. Values of p<0.05 were considered statistically significant for each analysis.

RESULTS

The DS system showed a significantly better score in the left lobe of the liver, and higher lesion-to-liver and spleen-to-liver CRs in the left and right lobes of the liver, compared with SSBTC+ or SSBTC- (p<0.05). The DS system and SSBTC- showed no significant differences in scores in the right lobe of the liver but they showed significantly better scores than SSBTC+ (p<0.05).

CONCLUSION

The DS system is more advantageous to improve the homogeneity of signal intensity distribution and tissue contrast of Gd-EOB-MRI than the post-processing correction method independently of the site.

摘要

目的

评估双源并行射频系统(DS 系统)在钆乙氧基苯甲基二乙三胺五乙酸增强磁共振成像(Gd-EOB-MRI)中的疗效。

材料与方法

26 例肝脏疾病患者分别采用 DS 系统或单源射频系统联合或不联合体调 CLEAR(一种用于校正 B1 不均匀性引起的信号丢失的后处理方法)行 Gd-EOB-MRI 肝胆期成像。分别对左、右肝叶进行评估。定性评估采用 3 分制评估信号强度分布均匀性评分。定量评估计算病变与肝脏、脾脏与肝脏的对比比(CR)。采用双向方差分析和 Tukey Honestly Significant Difference 事后检验比较三种技术之间的评分和 CR。每种分析中,p<0.05 被认为具有统计学意义。

结果

DS 系统在左肝叶的评分明显更好,左、右肝叶的病变与肝脏、脾脏与肝脏的 CR 更高,与 SSBTC+或 SSBTC-相比(p<0.05)。DS 系统和 SSBTC-在右肝叶的评分无显著差异,但评分明显优于 SSBTC+(p<0.05)。

结论

DS 系统在改善 Gd-EOB-MRI 信号强度分布和组织对比的均匀性方面比独立于部位的后处理校正方法更有优势。