Palmer R H, Aron D N
Department of Small Animal Medicine and Surgery, College of Veterinary Medicine, Texas A&M University, College Station 77843.
Vet Surg. 1990 Nov-Dec;19(6):440-5. doi: 10.1111/j.1532-950x.1990.tb01231.x.
Complications developed with the clinical use of Ellis pins in external skeletal fixation in seven dogs weighing 11 to 24 kg. Pins broke at the threaded-nonthreaded shaft junction in six dogs, and there was radiographic evidence of pin loosening six dogs. One pin pulled out causing loss of fixation in one dog. These complications with the recently introduced Ellis pin emphasize a need for further evaluation of its proper use.
在7只体重11至24千克的犬只中,使用埃利斯针进行外骨骼固定时出现了并发症。6只犬的针在螺纹与非螺纹轴交界处断裂,6只犬有影像学证据显示针松动。1只犬的一根针拔出导致固定失效。最近引入的埃利斯针出现的这些并发症强调了需要对其正确使用进行进一步评估。