Schulte W, Wagner M
Zentrum für Zahn-, Mund- und Kieferheilkunde der Universität Tubingen.
Dtsch Zahnarztl Z. 1990 Jul;45(7):394-9.
Using a serial periotest unit we determined the periotest values for disclusion, slight centric occlusion, maximum intercuspation and axially in 102 subjects with a total of 1769 teeth in periodontally healthy dentitions. All teeth were grouped according to the criteria presence or lack of antagonist contact in maximum intercuspation, excentric abrasion facets and occlusal restorations, and statistically evaluated. The difference-normal range for physiologic occlusal loads can be assessed by comparative periotest measurements in disclusion and maximum intercuspation. Thus we have, for the first time, a quantitative, objectifiable parameter for the assessment of the "dimension of occlusion", which at the same time provides information on the functional load on the periodontia of antagonists. Premolar teeth with occlusal restorations tended to have significantly greater differences in the periotest values than teeth with natural occlusal surfaces. By comparing these values with the normal range of differences in the periotest value, a high degree of precision can be achieved in adjusting occlusal restorations. Axial measurements revealed a markedly higher shock absorbing capacity of the periodontia than horizontal measurements.
我们使用连续牙周测试装置,测定了102名牙周健康牙列患者共1769颗牙齿在牙合分离、轻度正中牙合、最大牙尖交错位及轴向的牙周测试值。所有牙齿根据最大牙尖交错位时有无拮抗牙接触、非正中磨耗面及牙合面修复体的标准进行分组,并进行统计学评估。生理牙合负荷的差异-正常范围可通过比较牙合分离和最大牙尖交错位时的牙周测试测量值来评估。因此,我们首次获得了一个用于评估“牙合维度”的定量、可客观化的参数,该参数同时提供了关于拮抗牙牙周组织功能负荷的信息。有牙合面修复体的前磨牙牙周测试值的差异往往比天然牙合面的牙齿显著更大。通过将这些值与牙周测试值差异的正常范围进行比较,在调整牙合面修复体时可实现高度的精确性。轴向测量显示牙周组织的减震能力明显高于水平测量。