Institute of Forensic Medicine, Forensic Toxicology, Freiburg, Germany.
J Mass Spectrom. 2012 Jun;47(6):778-85. doi: 10.1002/jms.3021.
An assay based on liquid chromatography/tandem mass spectrometry is presented for the fast, precise and sensitive quantitation of Δ9-tetrahydrocannabinolic acid A (THCA) in serum. THCA is the biogenetic precursor of Δ9-tetrahydrocannabinol in cannabis and has aroused interest in the pharmacological and forensic field especially as a potential marker for recent cannabis use. After addition of deuterated THCA, synthesized from D(3)-THC as starting material, and protein precipitation, the analytes were separated using gradient elution on a Luna C18 column (150 × 2.0 mm × 5 µm) with 0.1% formic acid and acetonitrile/0.1% formic acid. Data acquisition was performed on a triple quadrupole linear ion trap mass spectrometer in multiple reaction monitoring mode with negative electrospray ionization. After optimization, the following sample preparation procedure was used: 200 μL serum was spiked with internal standard solution and methanol and then precipitated 'in fractions' with 500 μL ice-cold acetonitrile. After storage and centrifugation, the supernatant was evaporated and the residue redissolved in mobile phase. The assay was fully validated according to international guidelines including, for the first time, the assessment of matrix effects and stability experiments. Limit of detection was 0.1 ng/mL, and limit of quantification was 1.0 ng/mL. The method was found to be selective and proved to be linear over a range of 1.0 to 100 ng/mL using a 1/x weighted calibration model with regression coefficients >0.9996. Accuracy and precision data were within the required limits (RSD ≤ 8.6%, bias: 2.4 to 11.4%), extractive yield was greater than 84%. The analytes were stable in serum samples after three freeze/thaw cycles and storage at -20 °C for one month.
建立了一种基于液相色谱/串联质谱的方法,用于血清中Δ9-四氢大麻酸 A(THCA)的快速、准确和灵敏定量。THCA 是大麻中 Δ9-四氢大麻醇的生物合成前体,在药理学和法医学领域引起了关注,特别是作为最近大麻使用的潜在标志物。在加入氘代 THCA 后,从 D(3)-THC 作为起始原料合成,并进行蛋白质沉淀,然后在 Luna C18 柱(150×2.0mm×5μm)上进行梯度洗脱,用 0.1%甲酸和乙腈/0.1%甲酸进行洗脱。在三重四极杆线性离子阱质谱仪上以负离子电喷雾电离的多反应监测模式进行数据采集。经过优化,采用以下样品制备程序:200μL 血清加入内标溶液和甲醇,然后用 500μL 冰冷的乙腈“分份”沉淀。沉淀后储存和离心,上清液蒸发,残渣用流动相重新溶解。该方法按照国际指南进行了全面验证,包括首次评估基质效应和稳定性实验。检测限为 0.1ng/mL,定量限为 1.0ng/mL。该方法具有选择性,并通过 1/x 加权校准模型证明在 1.0 至 100ng/mL 的范围内呈线性,回归系数>0.9996。准确度和精密度数据在要求的范围内(RSD≤8.6%,偏差:2.4 至 11.4%),提取回收率大于 84%。在经过三个冻融循环和在-20°C 下储存一个月后,血清样品中的分析物稳定。