开发一种压力测量装置以优化淋巴水肿的压迫治疗,并评估模拟磨损情况下服装压力的变化。

Development of a pressure-measuring device to optimize compression treatment of lymphedema and evaluation of change in garment pressure with simulated wear and tear.

作者信息

Brorson Håkan, Hansson Emma, Jense Erik, Freccero Carolin

机构信息

Department of Clinical Sciences, Lund University, Plastic and Reconstructive Surgery, Skåne University Hospital, Malmö, Sweden.

出版信息

Lymphat Res Biol. 2012 Jun;10(2):74-80. doi: 10.1089/lrb.2012.0003.

Abstract

The use of compression garments in treating lymphedema following treatment of genital (penis, testes, uterus, cervical) and breast cancer treatment is a well-established practice. Although compression garments are classified in compression classes, little is known about the actual subgarment pressure exerted along the extremity. The aims of this study were to establish an in vitro method for measuring subgarment pressure along the extremity and to analyze initial and over time subgarment pressure of compression garments from three manufacturers. The measurements were performed with I-scan(®) (Tekscan Inc.) pressure measuring equipment once a week during a period of 4 weeks. Wear and tear was simulated by washing and putting on the garments on plastic legs every day. There was a statistically significant difference between the garments of some of manufacturers. There was no difference between garments from the same manufacturer. No significant decrease of subgarment pressure was observed during the trial period. The study demonstrated that Tekscan pressure-measuring equipment could measure subgarment pressure in vitro. The results may indicate that there was a difference in subgarment pressure exerted by garments from different manufacturers and that there was no clear decrease in subgarment pressure during the first four weeks of usage.

摘要

在生殖器(阴茎、睾丸、子宫、宫颈)和乳腺癌治疗后使用压力衣治疗淋巴水肿是一种成熟的做法。尽管压力衣按压力等级分类,但对于沿肢体施加的实际衣下压力却知之甚少。本研究的目的是建立一种体外测量沿肢体衣下压力的方法,并分析来自三个制造商的压力衣的初始和随时间变化的衣下压力。在4周的时间里,每周使用I-scan®(Tekscan公司)压力测量设备进行一次测量。通过每天在塑料腿上洗涤和穿上压力衣来模拟磨损。一些制造商的压力衣之间存在统计学上的显著差异。同一制造商的压力衣之间没有差异。在试验期间未观察到衣下压力有显著下降。该研究表明,Tekscan压力测量设备可以在体外测量衣下压力。结果可能表明,不同制造商的压力衣施加的衣下压力存在差异,并且在使用的前四周衣下压力没有明显下降。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验