Institute of Organic Chemistry, Polish Academy of Sciences, Kasprzaka 44/52, 01-224 Warsaw, Poland.
Rapid Commun Mass Spectrom. 2012 Aug 15;26(15):1705-10. doi: 10.1002/rcm.6280.
The aim of this study was to determine whether Atmospheric Pressure Photoionization (APPI) was better suited for the mass spectrometric (MS) analysis of polyisoprenoid alcohols than the commonly used Electrospray Ionization (ESI) method. The APPI method should make possible the use of non-polar solvents without any of the additives required by ESI, together with improved detection limits.
The liquid chromatography (LC)/APPI-MS and LC/ESI-MS spectra of polyisoprenoid alcohol standards were acquired in both positive and negative ion mode, in methanol and hexane, using a triple quadrupole/linear ion trap tandem mass spectrometer equipped with both an ESI and an APPI ion source.
In the positive ion mode peaks corresponding to M + H - H(2)O and M + H ions were observed in the APPI-MS spectra of polyprenols and dolichols, respectively. In the negative ion mode peaks corresponding to M + O(2) and M + Cl ions were observed for both classes of polyisoprenoid alcohols. The detection limit of polyisoprenoid alcohols was established at the level of 10 pg.
APPI turned out to be a method of choice for the identification and quantitation of polyisoprenoid alcohols by MS using both polar and non-polar solvents. APPI also enabled greater differentiation of polyprenols and dolichols occurring together in natural samples and gave much better TIC chromatograms without the need for the post-column salt addition required by ESI.
本研究旨在确定大气压光电离(APPI)是否比常用的电喷雾电离(ESI)方法更适合于聚异戊二烯醇的质谱(MS)分析。APPI 方法应该可以在不使用 ESI 所需的任何添加剂的情况下使用非极性溶剂,并提高检测限。
在正离子模式和负离子模式下,使用配备有 ESI 和 APPI 离子源的三重四极杆/线性离子阱串联质谱仪,在甲醇和己烷中分别获得聚异戊二烯醇标准品的液相色谱(LC)/APPI-MS 和 LC/ESI-MS 谱。
在正离子模式下,分别观察到聚异戊烯醇和多萜醇的 APPI-MS 谱中出现[M+H-H2O]+和[M+H]+离子的峰。在负离子模式下,两类聚异戊二烯醇均观察到[M+O2]-•和[M+Cl]-离子的峰。聚异戊二烯醇的检测限设定在 10pg 水平。
APPI 被证明是一种通过 MS 对聚异戊二烯醇进行鉴定和定量的首选方法,既可以使用极性溶剂,也可以使用非极性溶剂。APPI 还能够更好地区分在天然样品中同时存在的聚异戊烯醇和多萜醇,并提供更好的总离子流色谱图,而无需 ESI 所需的柱后加盐。