Myocardial first-pass perfusion imaging with hybrid-EPI: frequency-offsets and potential artefacts.

机构信息

Faculty of Medicine, Imperial College, London, UK.

出版信息

J Cardiovasc Magn Reson. 2012 Jun 25;14(1):44. doi: 10.1186/1532-429X-14-44.

Abstract

BACKGROUND

First-pass myocardial perfusion is often imaged with a tailored hybrid centric interleaved echo-planar-imaging sequence, providing rapid image acquisition with good contrast enhancement. The centric interleaved phase-encode order minimises the effective time-of-echo but it is sensitive to frequency-offsets. This short article aims to show possible artefacts that might originate with this sequence, in the context of first-pass perfusion imaging, when frequency-offsets are present. Non-uniform magnitude modulation effects were also analysed.

METHODS

Numerical and phantom simulations were used to illustrate the effects of frequency-offsets and non-uniform magnitude modulation with this sequence in a typical perfusion protocol. In vivo data was post-processed to analyse the h-EPI's sensitivity to the frequency-offsets.

RESULTS

The centric phase-order was shown to be highly sensitive to frequency-offsets due to its symmetrical phase slope. Resulting artefacts include blurring, and splitting of the image into two identical copies along the phase-encode direction. It was also shown that frequency-offsets can introduce signal loss and ghosting of the right ventricle signal into the myocardium. The in vivo results were confirmed by numerical and phantom simulations. Magnitude modulation effects were found to be small.

CONCLUSIONS

Imaging first-pass myocardial perfusion with an hybrid centric echo-planar-imaging sequence can be corrupted with ghosting and splitting of the image due to frequency-offsets.

摘要

背景

首过心肌灌注通常采用定制的混合中心交错式 echo-planar-imaging 序列进行成像,该序列提供快速的图像采集和良好的对比度增强。这种中心交错相位编码顺序最小化了有效回波时间,但对频率偏移很敏感。本文旨在展示在存在频率偏移的情况下,这种序列在首过灌注成像中可能产生的伪影。还分析了非均匀幅度调制效应。

方法

使用数值和体模模拟来说明在典型灌注方案中,频率偏移和非均匀幅度调制对该序列的影响。对体内数据进行后处理以分析 h-EPI 对频率偏移的敏感性。

结果

由于其对称的相位斜率,中心相位顺序对频率偏移非常敏感。由此产生的伪影包括图像沿相位编码方向模糊和分裂成两个相同的副本。还表明,频率偏移会导致右心室信号的信号丢失和鬼影进入心肌。体内结果得到了数值和体模模拟的证实。发现幅度调制效应很小。

结论

使用混合中心 echo-planar-imaging 序列进行首过心肌灌注成像时,由于频率偏移,图像可能会出现鬼影和分裂。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/cca1/3457847/0b53a0e316d0/1532-429X-14-44-1.jpg

相似文献

1
Myocardial first-pass perfusion imaging with hybrid-EPI: frequency-offsets and potential artefacts.
J Cardiovasc Magn Reson. 2012 Jun 25;14(1):44. doi: 10.1186/1532-429X-14-44.
2
Quantification and reduction of ghosting artifacts in interleaved echo-planar imaging.
Magn Reson Med. 1997 Sep;38(3):429-39. doi: 10.1002/mrm.1910380312.
5
A method of generalized projections (MGP) ghost correction algorithm for interleaved EPI.
IEEE Trans Med Imaging. 2004 Jul;23(7):839-48. doi: 10.1109/TMI.2004.827970.
6
Assessment of myocardial perfusion by magnetic resonance imaging.
Herz. 1997 Feb;22(1):16-28. doi: 10.1007/BF03044567.
7
Propeller EPI in the other direction.
Magn Reson Med. 2006 Jun;55(6):1298-307. doi: 10.1002/mrm.20890.
8
View angle tilting echo planar imaging for distortion correction.
Magn Reson Med. 2012 Oct;68(4):1211-9. doi: 10.1002/mrm.23320. Epub 2011 Dec 28.
9
Analysis of spatiotemporal fidelity in quantitative 3D first-pass perfusion cardiovascular magnetic resonance.
J Cardiovasc Magn Reson. 2017 Jan 27;19(1):11. doi: 10.1186/s12968-017-0324-z.
10
Centric ordering is superior to gradient moment nulling for motion artifact reduction in EPI.
J Magn Reson Imaging. 1997 Nov-Dec;7(6):1122-31. doi: 10.1002/jmri.1880070627.

引用本文的文献

1
MRI-Based Assessment of Brain Tumor Hypoxia: Correlation with Histology.
Cancers (Basel). 2023 Dec 27;16(1):138. doi: 10.3390/cancers16010138.
4
Realistic Dynamic Numerical Phantom for MRI of the Upper Vocal Tract.
J Imaging. 2020 Aug 27;6(9):86. doi: 10.3390/jimaging6090086.
5
A review of 3D first-pass, whole-heart, myocardial perfusion cardiovascular magnetic resonance.
J Cardiovasc Magn Reson. 2015 Aug 1;17(1):68. doi: 10.1186/s12968-015-0162-9.
6
Review of Journal of Cardiovascular Magnetic Resonance 2013.
J Cardiovasc Magn Reson. 2014 Dec 5;16:100. doi: 10.1186/s12968-014-0100-2.
8
Review of Journal of Cardiovascular Magnetic Resonance 2012.
J Cardiovasc Magn Reson. 2013 Sep 4;15(1):76. doi: 10.1186/1532-429X-15-76.
9
Cardiovascular magnetic resonance artefacts.
J Cardiovasc Magn Reson. 2013 May 22;15(1):41. doi: 10.1186/1532-429X-15-41.

本文引用的文献

4
Quantitative myocardial distribution volume from dynamic contrast-enhanced MRI.
Magn Reson Imaging. 2008 May;26(4):532-42. doi: 10.1016/j.mri.2007.10.003. Epub 2008 Feb 20.
6
T2* measurement during first-pass contrast-enhanced cardiac perfusion imaging.
Magn Reson Med. 2006 Nov;56(5):1132-4. doi: 10.1002/mrm.21061.
8
Magnetic resonance perfusion measurements for the noninvasive detection of coronary artery disease.
Circulation. 2003 Jul 29;108(4):432-7. doi: 10.1161/01.CIR.0000080915.35024.A9. Epub 2003 Jul 14.
9
Generalized autocalibrating partially parallel acquisitions (GRAPPA).
Magn Reson Med. 2002 Jun;47(6):1202-10. doi: 10.1002/mrm.10171.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验