Faculty of Medicine, Imperial College, London, UK.
J Cardiovasc Magn Reson. 2012 Jun 25;14(1):44. doi: 10.1186/1532-429X-14-44.
First-pass myocardial perfusion is often imaged with a tailored hybrid centric interleaved echo-planar-imaging sequence, providing rapid image acquisition with good contrast enhancement. The centric interleaved phase-encode order minimises the effective time-of-echo but it is sensitive to frequency-offsets. This short article aims to show possible artefacts that might originate with this sequence, in the context of first-pass perfusion imaging, when frequency-offsets are present. Non-uniform magnitude modulation effects were also analysed.
Numerical and phantom simulations were used to illustrate the effects of frequency-offsets and non-uniform magnitude modulation with this sequence in a typical perfusion protocol. In vivo data was post-processed to analyse the h-EPI's sensitivity to the frequency-offsets.
The centric phase-order was shown to be highly sensitive to frequency-offsets due to its symmetrical phase slope. Resulting artefacts include blurring, and splitting of the image into two identical copies along the phase-encode direction. It was also shown that frequency-offsets can introduce signal loss and ghosting of the right ventricle signal into the myocardium. The in vivo results were confirmed by numerical and phantom simulations. Magnitude modulation effects were found to be small.
Imaging first-pass myocardial perfusion with an hybrid centric echo-planar-imaging sequence can be corrupted with ghosting and splitting of the image due to frequency-offsets.
首过心肌灌注通常采用定制的混合中心交错式 echo-planar-imaging 序列进行成像,该序列提供快速的图像采集和良好的对比度增强。这种中心交错相位编码顺序最小化了有效回波时间,但对频率偏移很敏感。本文旨在展示在存在频率偏移的情况下,这种序列在首过灌注成像中可能产生的伪影。还分析了非均匀幅度调制效应。
使用数值和体模模拟来说明在典型灌注方案中,频率偏移和非均匀幅度调制对该序列的影响。对体内数据进行后处理以分析 h-EPI 对频率偏移的敏感性。
由于其对称的相位斜率,中心相位顺序对频率偏移非常敏感。由此产生的伪影包括图像沿相位编码方向模糊和分裂成两个相同的副本。还表明,频率偏移会导致右心室信号的信号丢失和鬼影进入心肌。体内结果得到了数值和体模模拟的证实。发现幅度调制效应很小。